Scout SPX-SJX-SKX FAQ
AppNotes.v20180726_ AG
For any further questions, please e-mail us at [email protected], or call us at 973-377-9000 #2 or ext. 7852 from 8AM to 5PM EST
Topic:
This list compiles frequently asked questions about the Scout SPX-SKX-SJX Series:
a) How do I calibrate my unit
b) How do I switch between units of measure
c) How do I turn on the units of measure
How to Calibrate
The Scout has two methods of calibration: Span and Linearity. Following is a quick guide on how to
perform a Span calibration. Span calibration uses two calibration points: zero and a specified calibration
weight (usually full scale capacity).
You can also click
to view a short video on this topic.
Before or during calibration:
a) Have the appropriate calibration masses ready
b) Clear any obstructions around the pan; make sure the transportation lock under the pan and the
security switch under the unit are unlocked
c) The unit is on for at least 5 minutes to warm up and stabilize
d) The unit is level
e) The location is stable: avoid vibrations, drafts, electromagnetic fields and extreme temperatures
You can also download the user manual
– refer to Section 3.3: Calibration Operation and the spec
table for available calibration mass requirements.
1. Press and hold [MENU] until MENU is displayed.
2. Release the button once MENU is displayed.
3. CAL is displayed - press [YES]
4. SPAN is displayed
– with nothing on the pan, press [YES]
--C-- flashes while the 0 value is stored
. . .