Scout SPX-SJX-SKX FAQ 

AppNotes.v20180726_ AG 

For any further questions, please e-mail us at [email protected], or call us at 973-377-9000 #2 or ext. 7852 from 8AM to 5PM EST 

Topic: 

This list compiles frequently asked questions about the Scout SPX-SKX-SJX Series: 

a)  How do I calibrate my unit 
b)  How do I switch between units of measure 
c)  How do I turn on the units of measure 

 

How to Calibrate 

The Scout has two methods of calibration: Span and Linearity.  Following is a quick guide on how to 
perform a Span calibration.  Span calibration uses two calibration points: zero and a specified calibration 
weight (usually full scale capacity).   

You can also clic

here

 to view a short video on this topic. 

Before or during calibration: 

a)  Have the appropriate calibration masses ready 
b)  Clear any obstructions around the pan; make sure the transportation lock under the pan and the 

security switch under the unit are unlocked 

c)  The unit is on for at least 5 minutes to warm up and stabilize 
d)  The unit is level 
e)  The location is stable: avoid vibrations, drafts, electromagnetic fields and extreme temperatures 

You can also download the user manual 

here

 

– refer to Section 3.3: Calibration Operation and the spec 

table for available calibration mass requirements.

 

1.  Press and hold [MENU] until MENU is displayed.  

 

2.  Release the button once MENU is displayed. 

 

 

 

3.  CAL is displayed - press [YES] 

 

 

 

4.  SPAN is displayed 

– with nothing on the pan, press [YES] 

 

 

 

--C-- flashes while the 0 value is stored 

 

 

. . . 