![OHAUS Scout Pro Series Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/ohaus/scout-pro-series/scout-pro-series_service-manual_1633114021.webp)
4-3
CHAPTER 4 TESTING
4.2.2 Off-Center Load Test
The Off-Center Load Test is used to determine whether displayed weight values are affected by
moving the sample to different areas of the Pan (6).
Test
1. Place 1/2 of the balance capacity in the
center
of the Pan (1).
2. Note the reading.
3. Move the mass halfway (between the center and the edge) to the front of the Pan (1). Note
any differences in the displayed weight reading.
4. Repeat this test for the back, left, and right positions of the Pan.
5. Maximum allowable change in displayed weight readings is + 2 count for each of the four
positions. If this reading is exceeded, it usually indicates a defective Transducer (8).
4.2.3 Linearity Test
This test is used to determine the linearity of the unit throughout its operating range. The masses used
to perform this test must be adjusted to ASTM Class 4 Tolerance, or better.
NOTE:
The balance must pass the Off-Center Load Test and Re-
peatability Test before the Linearity Test may be performed.
Test
Loads do not have to be test weights. They can be anything that totals the load value. The test mass
can be anything that weighs near the test mass value
TABLE 4-2. TEST MASSES
Capacity
Test mass
Load 1
Load 2
Load 3
Load 4
200g
1g
50g
100g
150g
190g
400g x 0.01
1g
100g
200g
300g
390g
400g x 0.1
5g
100g
200g
300g
400g
600g
5g
150g
300g
450g
600g
2000g
5g
500g
1000g
1500g
2000g
4000g
5g
1000g
2000g
3000g
4000g
6000g
10g
1500g
3000g
4500g
6000g
1. Place the test mass on the balance, record the weight and then remove.
2. Place Load 1 on the balance and press the
ON/ZERO/
Off
button.
3. Place the test mass on the balance, record the weight and then remove.
4. Place Load 2 on the balance and press the
ON/ZERO/
Off
button.
www.mirvesov.ru
Тел
.: (495) 921-44-57
Содержание Scout Pro Series
Страница 2: ...www mirvesov ru 495 921 44 57...
Страница 4: ...www mirvesov ru 495 921 44 57...
Страница 10: ...1 4 CHAPTER 1 INTRODUCTION www mirvesov ru 495 921 44 57...
Страница 24: ...4 6 CHAPTER 4 TESTING www mirvesov ru 495 921 44 57...
Страница 26: ...www mirvesov ru 495 921 44 57...
Страница 38: ...B 4 APPENDIX B FACTORY CALIBRATIONS www mirvesov ru 495 921 44 57...
Страница 39: ...www mirvesov ru 495 921 44 57...