![OHAUS RU-U2 Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/ohaus/ru-u2/ru-u2_service-manual_4734483011.webp)
Service Manual RU
OHAUS
Page 11 / 53
11/01/2010
Ohaus
3. Troubleshooting
Following is a list of symptoms that could occur, and the recommended action to
correct the problem.
SCALE INOPERATIVE/BLANK DISPLAYS
1. Check AC power at outlet.
2. Check Voltage on Main Logic PCB. If voltages are zero, check Power Supply.
3. If test point voltage is good replace Main Logic PCB.
4. Check Display Cable on Main Logic PCB and Operator Display PCB.
Check Display PCB.
KEYBOARD INOPERATIVE
1. Check keyboard connection on keyboard controller PCB.
2. Check Keyboard controller connection to Main Logic, ensure all connections are
secure.
3. If all connections are secure and keyboard is inoperative:
a) Replace keyboard controller PCB.
b) Check keyboard once again.
c) If Keyboard is still inoperative, replace keyboard.
SCALE WON'T ZERO (DISPLAY "PUT PLATTER ON" or "ZERO NOT CAPTURE")
1. Check platter and spider for obstructions. Turn power off then back on.
2. Recalibrate.
3. Check Load Cell supply voltage.
4. Replace Load Cell.
LOSING PROGRAMMED DATA
1. Check Error Report of Service Mode.
2. Check Main Logic voltage at test points.
3. Replace Main Logic PCB.
Содержание RU-U2
Страница 1: ...RU RU Service Manual RU RU U2 RU C2 Edition January 2010...
Страница 20: ...Service Manual RU OHAUS Page 20 53 11 01 2010 Ohaus...
Страница 25: ...Service Manual RU OHAUS Page 25 53 11 01 2010 Ohaus 11 Interconnecting Diagram...
Страница 29: ...Service Manual RU OHAUS Page 29 53 11 01 2010 Ohaus 15 Replace load cell...
Страница 33: ...Service Manual RU OHAUS Page 33 53 11 01 2010 Ohaus 17 Replace printer I...
Страница 49: ...Service Manual RU OHAUS Page 49 53 11 01 2010 Ohaus...