background image

EN-30                                                        Ranger™ Count 3000

 

Series

 

 

 

7. TECHNICAL DATA  

The technical data is valid under the following ambient conditions: 

Indoor use only 

Operating temperature: -10°C to 40°C 

Relative humidity: 20% to 85% relative humidity, non-condensing 

Height above sea level: Up to 2000 m 

Power: AC power 100-240V 50/60 Hz, internal rechargeable sealed lead-acid battery 

Current rating: 0.5A 

Mains supply voltage fluctuations: up to ± 10% of the nominal voltage 

Installation category: II 

Pollution degree: 2 

 

7.1 Specifications

 

TABLE 7-1 

MODEL 

RC31P1502 

RC31P3 

RC31P6 

RC31P15 

RC31P30 

Capacity x Readability 

(Max x d non-approved) 

1.5 kg x 0.00005 kg 

1500 g x 0.05 g 

3 lb x 0.0001 lb 

48 oz x 0.002 oz 

3 kg x 0.0001 kg 

3000 g x 0.1 g 

6 lb x 0.0002 lb 

96 oz x 0.005 oz 

6 kg x 0.0002 kg 

6000 g x 0.2 g  

15 lb x 0.0005 lb 

240 oz x 0.01 oz 

15 kg x 0.0005 kg 

15000 g x 0.5 g 

30 lb x 0.001 lb 

480 oz x 0.02 oz 

30 kg x 0.001 kg 

30000 g x 1 g 

60 lb x 0.002 lb 

960 oz x 0.05 oz 

Maximum Displayed 

Resolution 

1:30000

 

Minimum Recommended 

Sample Weight 

1g / 0.002lb 

2g / 0.004lb 

4g / 0.01lb 

10g / 0.02lb 

20g / 0.04lb 

Minimum Recommended 

APW 

0.005g / 0.00001lb  0.01g / 0.00002lb  0.02g / 0.00005lb  0.05g / 0.0001lb 

0.1 g / 0.0002lb 

Capacity x Readability 

(Max x e approved) 

1.5 kg x 0.0005 kg 

1500 g x 0.5 g 

3 lb x 0.001 lb 

48 oz x 0.02 oz 

3 kg x 0.001 kg 

3000 g x 1 g 

6 lb x 0.002 lb 

96 oz x 0.05 oz 

6 kg x 0.002 kg 

6000 g x 2 g  

15 lb x 0.005 lb 

240 oz x 0.1 oz 

15 kg x 0.005 kg 

15000 g x 5 g 

30 lb x 0.01 lb 

480 oz x 0.2 oz 

30 kg x 0.01 kg 

30000 g x 10 g 

60 lb x 0.02 lb 

960 oz x 0.5 oz 

Approved Resolution 

1:3000 

Repeatability 

0.0001 kg 

0.0002 kg 

0.0005 kg 

0.001 kg 

0.002 kg 

Linearity 

0.0001 kg 

0.0002 kg 

0.0005 kg 

0.001 kg 

0.002 kg 

Weighing Units 

LFT OFF: g, kg, lb, oz, lb:oz 

LFT ON: g, kg, lb, oz 

EC and OIML Approved models: g, kg 

Tare Range 

To capacity by subtraction 

Stabilization Time 

≤ 1 second 

Safe Overload Capacity 

150 % of capacity 

Weight Display 

3 LCD with white LED backlight 

6-digit 7-segment, 20.5 mm / 0.8 in characters 

Keyboard 

20 mechanical buttons 

Battery Operating Time  

(at 20ºC) 

210 hours with backlight turned off 

Construction 

ABS plastic housing with 304 stainless steel platform 

Approval Class 

III 

Pan Dimensions 

225 x 300 mm / 8.9 x 11.8 in 

Shipping Dimensions 

397 x 447 x 195 mm / 15.6 x 17.6 x 7.7 in 

Net Weight  

4.2 kg / 9.3 lb 

Gross Weight  

5.45 kg / 12 lb 

 

 

 
 

 

 

 
 

 

 

 
 

Содержание RC31P1502-M

Страница 1: ...Ranger Count 3000 Series Instruction Manual...

Страница 2: ......

Страница 3: ...ty Precautions Please follow these safety precautions Verify that the AC input voltage printed on the data label matches the local AC power supply Do not drop loads on the platform Make sure that the...

Страница 4: ...firm steady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes Allow sufficient space around the scale 2 4 Leveling the Equipment The Ranger Count...

Страница 5: ...eration the battery symbol indicates the battery charge level The scale will automatically turn off when the batteries are fully discharged For maximum operating time the battery should be charged at...

Страница 6: ...s2 Off Turn scale off Units Change the weighing unit Switch between Check Weigh Check Count and Off Menu Enter user menu View the Audit Trail event counters extended press Menu Function Short Press Ye...

Страница 7: ...Zero symbol 11 Auto Optimization Symbol 3 Negative symbol 12 Dynamic symbol not used 4 Stable weight symbol 13 Battery charge symbol 5 Preset Tare Tare symbols 14 Pound Ounce Pound Ounce symbols 6 Poi...

Страница 8: ...ed in Checkweigh mode section 3 5 and will light up according to the following rules Red Loads Upper limit Green Loads Lower limit and Upper limit Yellow Loads Lower limit Figure 3 3 Below view of Ran...

Страница 9: ...0 Positive Sampling 1 Place the sample on the pan 2 Input the sample size with the numeric keypad The value is displayed in the count window Alternatively the Sample key can be used instead Pressing...

Страница 10: ...o display the statistical information To clear the accumulation data press the Tare key while the statistical information is displayed The display shows Clr aCC Press the Yes key to clear the stored d...

Страница 11: ...under the target pieces range the yellow LED will light If the sample is within the target pieces range the green LED will light If the sample is over the target pieces range the red LED will light P...

Страница 12: ...y reCall and lib are displayed in the weight and piece weight windows respectively 2 Long press the ID key edit and lib are momentarily displayed in the weight and piece weight windows respectively 3...

Страница 13: ...nd P r i n t 1 C O M 1 L O C k E n d Menu Items Reset Stable A Print Contnt Layout Data Tr End Reset Baud Parity Stop Handsh Alt Cm End L Cal L Setup L Read L Unit L Print L COM End Notes 1 Only visib...

Страница 14: ...setting and advances to the next item No Skips by the displayed menu Rejects the displayed setting or menu item and advances to the next available item Back Moves backwards through the upper and middl...

Страница 15: ...the span calibration 0 kg will be displayed Press Yes to accept C will be displayed while zero reading is stored Next the display shows the calibration weight value Place the specified calibration ma...

Страница 16: ...tiates a 2nd scale linearity calibration procedure GEO GEO Geographical Adjustment Factor GEO is used to adjust the calibration based on the current location Settings from 0 to 31 are available with 1...

Страница 17: ...layed at startup AUTO last unit in use when turned off kg kilograms g grams lb pounds oz ounces lb oz pound ounces Zero Range ZErO Set the percentage of scale capacity that may be zeroed 2 zero up to...

Страница 18: ...is disabled ACCEPT the beeper will sound when the weight is within the Accept range UNDER the beeper will sound when the weight is below the Under setting OVER the beeper will sound when the weight is...

Страница 19: ...ill maintain zero until a change of 0 5 divisions per second has been exceeded 1d the display will maintain zero until a change of 1 divisions per second has been exceeded 3d the display will maintain...

Страница 20: ...t 3000 Series 4 5 Unit Menu This sub menu activates units so they will be accessible with the Units button The units in the menu must be turned on to be active Note Available units vary by model and l...

Страница 21: ...printing occurs at the defined interval CONT printing occurs continuously ACCEPT printing occurs each time the display is within the Checkweigh accept range and stability criteria are met Content CON...

Страница 22: ...is printed Header Set the status OFF disabled ON the Header is printed Note See section 7 3 1 for how to enter Header line Footer Set the status OFF disabled ON the Footer is printed Note See section...

Страница 23: ...paper up four lines after printing FORM a form feed is appended to the printout Data Transfer data tr Output weighing results directly to a PC application OFF disabled ON enabled Notes Data Transfer F...

Страница 24: ...ps 19200 19200 bps Parity paritY Set the data bits and parity 7 EVEN 7 data bits even parity 7 ODD 7 data bits odd parity 7 NONE 7 data bits no parity 8 NONE 8 data bits no parity Stop bit stOP Set th...

Страница 25: ...ch can lock Menu settings to prevent tampering Default settings are in bold Reset reset Reset the Lock menu to factory defaults NO no reset YES reset Lock Cal l CAL Set the status OFF Calibration menu...

Страница 26: ...n to the current application mode 4 10 Additional Features To use the weigh below feature remove AC power from the scale and remove the protective cover for the weigh below opening Install a suitable...

Страница 27: ...2 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15 15 14 14 13 13 12 44 32 45 38 18 17 17 16 16 15 15 14 14 13 13 45 38 46 45 18 18 17 17 16 16 15 15 14 14 13 46 45 47 51 19...

Страница 28: ...lained in Section 4 2 3 Set the switch to Locked See figure 5 2 The Menu Lock switch limits changes to the Cal Setup Readout Unit and Print menus The switch in type approved models may set some scale...

Страница 29: ...he following menu settings are changed Dual Scale Capacity 2 Graduation 2 Zero Range Stable Range Auto Zero Tracking AZT Units kg g oz lb or lb oz or Stable Printing Dual Scale Capacity 2 Graduation 2...

Страница 30: ...Err 8 1 Power On Error Weight reading exceeds Power On Zero limit Err 8 2 Power On Error Weight reading below Power On Zero limit Err 8 3 Over Range Error Weight reading exceeds Overload limit Err 8 4...

Страница 31: ...1 800 526 0659 between 8 00 AM and 5 00 PM EST An OHAUS product service specialist will be available to provide assistance Outside the USA please visit our web site to locate the OHAUS office nearest...

Страница 32: ...20g 0 04lb Minimum Recommended APW 0 005g 0 00001lb 0 01g 0 00002lb 0 02g 0 00005lb 0 05g 0 0001lb 0 1 g 0 0002lb Capacity x Readability Max x e approved 1 5 kg x 0 0005 kg 1500 g x 0 5 g 3 lb x 0 00...

Страница 33: ...Ranger Count 3000 Series EN 31 7 2 Drawings Figure 7 1 Dimensions 311 mm 12 2 in 327 mm 12 9 in 117 mm 4 6 in...

Страница 34: ...ame as pressing Tare Key xT Download Tare value in grams positive values only Sending 0T clears tare if allowed PU Print current unit g kg lb oz lb oz xU Set scale to unit x 1 g 2 kg 3 lb 4 oz 5 lb oz...

Страница 35: ...eight string print format Check Weighing application Field Weight Right aligned Space Unit Right aligned Space Stability Space T N G PT Right aligned Space Application Status Right aligned Term Length...

Страница 36: ...descriptive data plate They may be put into service immediately Weighing Instruments to be verified in two stages have no green M metrology on the descriptive data plate and bear one of the preceding...

Страница 37: ...ns in Europe are available online Thank you for your contribution to environmental protection FCC Note This equipment has been tested and found to comply with the limits for a Class B digital device p...

Страница 38: ...has been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other th...

Отзывы: