PR Series Balance 

 

5.3.11 Approved Mode 

Use this menu to set the Legal for Trade status.    
 

OFF = standard operation. 

ON   = operation complies with Legal Metrology regulations.

    

 

 

Note:

 When Approved Mode is set to On, the menu settings are affected as follows:  

 

Calibration Menu: 

x

  For InCal models, only Internal Calibration is available. All other functions are hidden. 

x

  For ExCal models, the entire Calibration menu is hidden. 

 

Balance Setup Menu: 

x

  Filter Level is locked at the current setting.   

x

  Auto Zero Tracking is limited to 0.5 Division and Off. The selected setting is locked. 

x

  Auto Tare is locked at current setting. 

x

  Graduations are forced to 1 Division and the menu item is hidden. 

 

Communication Menu (Communication->Print Settings->Print Output): 

x

  Stable Weight Only is locked On. 

x

  Numeric Value Only is locked Off. 

 
Communication Menu (Communication->Print Settings->Auto Print): 

x

  Auto print mode selections are limited to Off, On Stability, and Interval. Continuous is not available. 

 

Lockout Menu: 

x

  Menu is hidden 

 

 
 
 
 

5.4  Weighing Units

 

Enter this sub-menu to activate the desired units of measure.  
 
PR balances provide a choice of 10 units, which are all set On by default. 

Note:

 

Due to national laws, the balance may not include some of the units of measure listed below. 

 

Unit 

Display 

Gram 

Kilogram 

kg 

Milligram 

mg 

Carat 

ct 

Pound 

lb 

Ounce 

oz 

Ounce Troy 

ozt 

Penny weight 

dwt 

Newton 

Grain 

GN 

 
 
 
 
 
 
 
 

LFt 

Содержание PR124

Страница 1: ...OHAUS Pioneer PR Series User Guide...

Страница 2: ......

Страница 3: ...3 5 1 Menu Navigation EN 13 5 1 1 Changing Settings EN 13 5 2 Calibration EN 14 5 2 1 Calibration Sub menu InCal models EN 14 5 2 2 Internal Calibration not applicable to ExCal models EN 14 5 2 3 Cal...

Страница 4: ...27 6 1 Settings EN 27 6 2 Verification EN 27 6 3 Sealing EN 27 7 PRINTING EN 28 7 1 Connecting configuring and Testing the Printer Computer Interface EN 28 7 2 Output format EN 29 7 3 Printout Exampl...

Страница 5: ...eful information about the product Warning Symbols General Hazard Electrical Shock Hazard 1 4 Safety Precautions CAUTION Read all safety warnings before installing making connections or servicing this...

Страница 6: ...e is level before it is used or after its location is changed The PR balance has a level bubble in a small round window beside the display To level the balance adjust the 2 leveling feet until the bub...

Страница 7: ...tion When the PR balance is first installed or when it is moved to another location it must be calibrated to ensure accurate weighing results PR balances are classified into two categories InCal model...

Страница 8: ...re operation Secondary Function Press and Hold Off x Zeroing current value Unit x Changes weighing units Mode x Changes application mode Menu Cal x Enters the main menu Calibration is the first sub me...

Страница 9: ...its net weight is displayed After the container and the objects are removed the load will be displayed as a negative number Press Tare to clear Zero Press Zero to zero the balance 3 3 Overview of Part...

Страница 10: ...ill appear 4 The resulting value is displayed in the active unit of measure Item Settings To view or adjust the current settings Weighing Units Change the displayed unit See Section 5 4 for the detail...

Страница 11: ...ue to 15 please press No Then Put and 15 will flash simultaneously 4 Place 15 samples on the pan Press the Function Mode button so that the weight of the 15 samples is used to establish the average pi...

Страница 12: ...ly as parts are added APW Optimization occurs only when the number of pieces added to the pan is between one and three times the number already on the pan Print Settings Changing printing setup See Se...

Страница 13: ...is stored in memory The display will show 100 6 Remove the reference sample and place the test object on the pan The ratio of the test object to the reference sample weight is displayed as a percentag...

Страница 14: ...eled and calibrated The PR balance is equipped with a weigh below hook for weighing below the balance as shown below Before turning the balance over remove the pan and draft shield elements if present...

Страница 15: ...and Time Project Name Reset Time Format Ounce Troy Balance ID Time Setting Penny Weight Balance Name Brightness Newton User Name Auto Dim Grain Project Name LFT Application Name Result Gross Weight Ne...

Страница 16: ...span calibration value on the pan and note the difference in divisions between the nominal mass value and the actual balance reading If the difference is within division calibration adjustment is not...

Страница 17: ...e weights and 100 000 g place weight s of 100 g on the pan for calibration 5 Once the span calibration is completed successfully CAL Done will display Press any button to return to the previous screen...

Страница 18: ...calibration mass value will be shown in the display After the display shows 100 0000 g please place weight s of 100 g on the pan for calibration 5 Remove the weight of 100 g from the pan After a whil...

Страница 19: ...5 3 2 AZT Auto Zero Tracking Set the automatic zero tracking functionality Auto Zero Tracking Off disabled 0 5d display maintains zero up to a drift of 0 5 graduation per second 1d display maintains...

Страница 20: ...e Format Set the current date format Date Format YY MM DD MM DD YY DD MM YY Date Format MM DD YYYY DD MM YYYY YYYY MM DD 5 3 6 Date Setup Set the current date in the desired date format To set the cur...

Страница 21: ...m High Low Brightness Medium High Low 5 3 10 Auto Dim Set whether the balance automatically turns off the display backlight of the display Off disabled 10 minutes become dim if there is no motion for...

Страница 22: ...ked at current setting x Graduations are forced to 1 Division and the menu item is hidden Communication Menu Communication Print Settings Print Output x Stable Weight Only is locked On x Numeric Value...

Страница 23: ...ss No to advance to the next unit 3 Press Yes to set the unit displayed to the weighing unit 5 5 RS232 Interface Setup Enter this sub menu to customize RS232 standard settings Data may be output to ei...

Страница 24: ...data bits odd parity stop bit 1 7 O 2 7 data bits odd parity stop bit 2 Parity 8 data bits stop bit 1 5 5 3 Handshake Set the flow control method NONE no handshaking XON XOFF XON XOFF handshaking HARD...

Страница 25: ...2 On Stability printing occurs when the stability criteria are met When On Stability is selected set the conditions for printing Load Prints when the displayed load is stable Load and Zero Prints when...

Страница 26: ...e Name On the balance name is printed Off the balance name is not printed 5 6 9 User Name On the user name is printed Off the user name is not printed 5 6 10 Project Name On the project name is printe...

Страница 27: ...g 4 Lines move the paper up four lines after printing 1 Line 4 Lines 5 7 GLP Enter this menu to set the Good Laboratory Practices GLP 5 7 1 Header Enables the printing of GLP headings There are up to...

Страница 28: ...available for each Header setting The default is blank 5 8 Factory Reset Use this sub menu to reset the all menu settings to their Factory default settings Reset All resets all menus to their factory...

Страница 29: ...ode to On in the Balance Setup menu 5 As shown in Figure 6 1 press the push button inside the hole Note When Approved Mode is set to On external calibration can t be performed 6 2 Verification A weigh...

Страница 30: ...ted Send line ends Echo typed characters Wrap lines Use RS232 Interface Commands Section 9 6 1 to control the balance from a PC SPDC Software The Serial Port Data Collection SPDC software is provided...

Страница 31: ...haracters field contains CRLF Four CRLF or Form Feed ASCII 12 depending on the LINE FEED menu setting 8 When Numeric Only is set On only the Weight Field is printed left aligned 7 3 Printout Examples...

Страница 32: ...erified By ____________ OHAUS 07 26 2017 05 16 37 Balance ID Balance Name PR2202 User Name Project Name Span Calibration Calibration is done Reference weight 2000 00 g Actual weight 2000 22 g Differen...

Страница 33: ...ial glass cleaner Attention Do not use solvents harsh chemicals ammonia or abrasive cleaning agents 8 3 Troubleshooting TABLE 8 1 TROUBLESHOOTING Symptom Display Possible Cause Err 8 1 Over init zero...

Страница 34: ...ing linearly to 50 relative humidity at 40 C x Operability is assured at ambient temperatures between 5 C and 40 C x Mains supply voltage fluctuations up to 10 of the nominal voltage x Installation ca...

Страница 35: ...m 4 7 in 12 cm 4 7 in 12 cm 4 7 in 12 cm Span Calibration Points g 30 60 50 100 100 200 100 200 200 400 250 500 Linearity Calibration Points g 20 40 60 50 75 100 50 100 150 200 50 100 150 200 100 200...

Страница 36: ...7 1 in 18 cm 7 1 in 18 cm 7 1 in 18 cm 7 1 in 18 cm 7 1 in 18 cm Span Calibration Points g 750 1500 1000 2000 2000 4000 2000 4000 3000 6000 Linearity Calibration Points g 500 1000 1500 500 1000 1500...

Страница 37: ...nt weighing Platform Size diameter 3 5 in 9 cm 3 5 in 9 cm 4 7 in 12 cm 4 7 in 12 cm Span Calibration Points g 50 100 100 200 150 300 250 500 Linearity Calibration Points g 50 75 100 50 100 150 200 10...

Страница 38: ...Percent weighing Platform Size diameter 7 1 in 18 cm Span Calibration Points g 400 800 1000 2000 2000 4000 2500 5000 1000 2000 2000 4000 3000 6000 Linearity Calibration Points g 200 400 600 800 500 1...

Страница 39: ...PR Series Balance EN 37 9 2 Drawings and Dimensions Fully assembled dimensions Figure 9 1 0 001 g 0 0001 g model Figure 9 2 0 1 g 0 01 g model 11 93...

Страница 40: ...stable non zero weight and stable zero reading to be printed xP Set Auto Print to Interval Print x print interval 1 3600 sec 0P disable the interval Print 0P 0P disable interval print continuous prin...

Страница 41: ...ounting APW x in grams must have APW stored P Print Counting application APW x Set Percent application reference weight x in grams must have reference weight stored P Print Percent application referen...

Страница 42: ...le Operating mode Full duplex Transmission mode Bit serial asynchronous Transmission code ASCII Baud rates 1200 2400 4800 9600 19200 38400 firmware selectable Bits parity 7 bit even 7 bit odd 7 bit no...

Страница 43: ...he applicable weights and measures authorities If national regulations limit the validity period of the verification the user of the weighing instrument must strictly observe the re verification perio...

Страница 44: ...been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than...

Страница 45: ......

Страница 46: ...ices worldwide Con oficinas en todo el mundo Avec des bureaux partout dans le monde Mit B ros weltweit Con uffici in tutto il mondo www ohaus com 30372570 P N 30372570 E 2018 Ohaus Corporation all rig...

Отзывы: