3;6HULHV%DODQFH
(1
1RWH
ExCal models only have Span Calibration and Linearity Calibration.
,QWHUQDO&DOLEUDWLRQQRWDSSOLFDEOHWR([&DOPRGHOV
Calibration is accomplished with the internal calibration mass. Internal Calibration can be performed at any time,
provided the balance has warmed up to operating temperature and is level.
With the Balance turned ON and no load on the pan, the internal calibration can be performed.
Alternatively, press the
7DUH0HQX&DO
button and select
Internal Cal
to initiate the internal calibration.
The screen shows the status, and then press any button to return to the current application after calibration.
,Q&DO$GMXVWQRWDSSOLFDEOHWR([&DOPRGHOV
Use this calibration method to fine tune the effect of the Internal Calibration.
Calibration Adjust may be used to adjust the result of the Internal Calibration by +100 divisions.
1RWH
Before making a calibration adjustment, perform an Internal Calibration. To verify whether an adjustment is
needed, place a test mass equal to the
VSDQ FDOLEUDWLRQ YDOXH
on the pan and note the difference (in divisions)
between the nominal mass value and the actual balance reading. If the difference is /- division, calibration
adjustment is not required. If the difference e/-1 division, calibration adjustment is recommended.
Example:
Expected weight reading:
200.000 (Test mass value)
Actual weight reading: 200.014
Difference in gram:
–
0.014
Difference in division:
–
14 (InCal Adjust value)
To perform a Calibration Adjustment, select InCal Adjustment from the list of Calibration Menu; enter the value
(positive or negative divisions) to match the difference noted earlier in the procedure.
Recalibrate using Internal Calibration. After calibration, place the test mass on the pan and verify that the mass
value now matches the displayed value. If not, repeat the procedure until Internal Calibration reading agrees with
the test mass.
Once completed, the balance stores the Adjustment value and the display returns to the current application.
6SDQ&DOLEUDWLRQ
Span calibration uses two calibration points, one at zero load and the other at specified full load (span). For
detailed calibration mass information, please see the specification tables in section 9.1.
With the balance turned ON and no load on the pan, Span Calibration can be performed. The best accuracy is
achieved using the mass closest to the full span value.
6WHSVIRUVSDQFDOLEUDWLRQ
1.
Press and hold the
7DUH 0HQX&DO
button, and the
Calibration Menu will display.
www.
GlobalScientificSupply
.com
GLOBAL SCIENTIFIC SUPPLY
The Laboratory Supply Company!