EN-3 

 

2.2 Initial Calibration 

For best results, the instrument's calibration should be checked prior to first use. If adjustment is needed, refer to 
the Calibration section of the instruction manual. 

 

Steps for calibration 

1.  Press and hold the 

Tare / Menu-Cal

 button, and the 

Calibration Menu will display.

   

 

2.  Press 

Yes

 to enter the Calibration Menu.  

 

3.  To change the calibration mode, press 

No

 until 

SpaN

 

(span calibration) is displayed. 

 

4.  The Calibration mass value will be shown in the screen. 

After the display shows 

200.0000

 g, place weight(s) of 

200 g on the pan for calibration. To change to the 
calibration point of half full capacity (e.g. 100 g), press 
the 

Function / Mode

 button. After the screen shows 

"Place weights" and "100.0000 g", place weight(s) of 
100 g on the pan for calibration. 

 

5.  Once  the  span  calibration  is  completed  successfully, 

CAL.Done 

will display.   

Press any button to return to the previous screen. 

    

 

 

6.  Remove  the  weight(s),  and  the  reading  will  be  set  to 

zero. 

 

 

Содержание Pioneer PJX Series

Страница 1: ...PJX Series Balances Quick Start Guide Quick_Start_Guide_PJX_V1 ...

Страница 2: ......

Страница 3: ...low sufficient space 1 2 Leveling Be sure the balance is level before it is used or after its location is changed The PJX balance has a level bubble in a small round window beside the display To level the balance adjust the 4 leveling feet until the bubble is centered in the circle Please refer to the below figure for leveling ...

Страница 4: ... Display Button Primary Function Short Press On Zero If the balance is Off turns on the balance If balance is On sets zero Print Sends the current displayed value to the serial interface Function Operation is dependent on the application mode Tare Performs tare operation Secondary Function Long Press Off Zeroing current value Unit Changes weighing units Mode Changes application mode Menu Cal Enter...

Страница 5: ...ess No until SpaN span calibration is displayed 4 The Calibration mass value will be shown in the screen After the display shows 200 0000 g place weight s of 200 g on the pan for calibration To change to the calibration point of half full capacity e g 100 g press the Function Mode button After the screen shows Place weights and 100 0000 g place weight s of 100 g on the pan for calibration 5 Once t...

Страница 6: ... cleaned with a commercial glass cleaner Attention Do not use solvents harsh chemicals ammonia or abrasive cleaning agents 4 2 Troubleshooting Symptom Display Possible Cause Err 8 1 Weight reading exceeds Power On Zero limit Err 8 2 Weight reading below Power On Zero limit Err 8 3 Weight reading exceeds overload limit Err 8 4 Weight reading below underload limit Ref Err Reference Weight too small ...

Страница 7: ......

Страница 8: ...us Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 1 973 377 9000 Fax 1 973 944 7177 With offices worldwide www ohaus com 30469440 P N 30469440 A 2018 Ohaus Corporation all rights reserved ...

Отзывы: