2.2 Initial Calibration
For best results, the instrument's calibration should be checked prior to first use. If adjustment is needed, refer to
the Calibration section of the instruction manual.
Steps for calibration
1. Press and hold the
Tare / Menu-Cal
button, and the
Calibration Menu will display.
2. Press
Yes
to enter the Calibration Menu.
3. To change the calibration mode, press
No
until
SpaN
(span calibration) is displayed.
4. The Calibration mass value will be shown in the screen.
After the display shows
200.0000
g, place weight(s) of
200 g on the pan for calibration. To change to the
calibration point of half full capacity (e.g. 100 g), press
the
Function / Mode
button. After the screen shows
"Place weights" and "100.0000 g", place weight(s) of
100 g on the pan for calibration.
5. Once the span calibration is completed successfully,
CAL.Done
will display.
Press any button to return to the previous screen.
6. Remove the weight(s), and the reading will be set to
zero.