CHAPTER 4 TESTING
Navigator Series Service Manual
25
Ohaus Corporation www.ohaus.com
4 TESTING
Before and after servicing a Navigator scale, an operational test and various performance tests should
be made to confirm that the scale meets specifications. Turn the scale on and allow it to warm up for
at least one hour before performing these tests.
NOTE:
Make sure the test area is free from drafts and that the scale rests on a level and vibration-free
surface.
4.1 TEST MASSES REQUIRED
The masses required to test the Ohaus Navigator scales must meet the requirements of ASTM
Class 4 or OIML F2 Tolerance. The mass values are listed in Table 4-1.
TABLE 4-1
CALIBRATION MASS VALUES
Model
Span
Calibration mass
Linearity
Calibration masses
NV212
200g
100g, 200g
NV311(N or M)
200g
200g, 300g
NV511
300g
300g, 500g
NV1101
500g
500g, 1000g
NV2101
1kg
1kg, 2kg
NV3100(N or M)
2kg
2kg, 3kg
NV4101
2kg
2kg, 4kg
NV5101
3kg
3kg, 5kg
NVL511
300g
300g, 500g
NVL1101
500g
500g, 1000g
NVL2101
1kg
1kg, 2kg
NVL5101
3kg
3kg, 5kg
NVL10000
5kg
5kg, 10kg
NVL20000
10kg
10kg, 20kg
NVT1601
1kg
1kg,1.5kg
NVT1601(N or M)
1kg
1kg,1.5kg
NVT3201
2kg
2kg, 3kg
NVT3200(N or M)
2kg
2kg, 3kg
NVT6400(N or M)
5kg
3kg, 6kg
NVT6401
5kg
3kg, 6kg
NVT10001
5kg
5kg, 10kg
NVT16000
10kg
10kg, 15kg
NVT16000(N or M)
10kg
10kg, 15kg