EN-8
Navigator
®
Err 8.1
ˉ
Load
ˉ
Power on zero range exceeded
Clear pan, check Shipping Lock setting
Err 8.2 _Load_
Power on zero under range
Install pan, check Shipping Lock setting
Err 8.3
ˉ
Load
ˉ
Overload (>cap+9e)
Load exceeds balance maximum capacity
Err 8.4 _Load_
Under load
Reading below min. range - Re-install pan.
Err 8.6 999999
Displayed value >999999
Result exceeds display capability.
Err 9 Data
Internal data error.
Contact an authorized service agent
Err 13 mMEmM
Fail to write EEPROM.
Contact an authorized service agent
Err 53 CSumM
Invalid checksum data
Contact an authorized service agent
5.2 Service Information
If the troubleshooting section does not resolve or describe your problem, contact your authorized
OHAUS service agent. For service assistance or technical support in the United States call toll-free 1-
800-526-0659 between 8:00 AM and 5:00 PM EST. An OHAUS product service specialist will be
available to provide assistance. Outside the USA, please visit our web site,
www.ohaus.com
to locate
the OHAUS office nearest you.
5.3 Accessories
Rechargeable Battery Kit
83032106 (NVL, NVT)
RS232 Interface Kit
83032107 (NV, NVL, NVT)
USB Interface Kit
83032108 (NV, NVL, NVT)
Ethernet Interface Kit
83032109 (NV, NVL, NVT)
In-Use Cover Kit
83032223 (NV)
83032222 (NVL) 83032221 (NVT)
Carrying Case Kit
83032226 (NV)
83032225 (NVL) 83032224 (NVT)
Printers and Cables
Contact OHAUS
6. TECHNICAL DATA
The technical data is valid under the following ambient conditions:
Ambient temperature: 10° C to 40° C, Approved models (M) 0° C to 40° C
Relative humidity: 20% to 85% relative humidity, non-condensing
Height above sea level: Up to 4000 m
Operability: assured at ambient temperatures between 0° C and 40° C
Power: AC Adapter – 12VDC 420mA output, 4 batteries or rechargeable battery
Protection: dust and water
Pollution degree: 2
Installation category: Class III
EMC: See Declaration of Conformity