EN-8 MB27
Troubleshooting
Symptom / Display
Possible Cause
Remedy
Cannot turn on
No power to analyzer
Verify connections, fuse and voltage
Weight value blinking Sample weight less than 0.5g
Increase sample size
“Tare” shown
Pan weight needs set to zero
Press Tare
“Close Cover” shown
Cover needs to be closed before starting test
Close cover
Poor accuracy
Improper calibration
Unstable environment
Perform calibration
Move the analyzer to suitable location
Cannot calibrate
Unstable environment
Incorrect calibration masses
Move the analyzer to suitable location
Use correct calibration masses
Err 7.0
Time out
Err 8.1
Pan support has load during power on
Remove weight from pan support
Err 8.2
Pan support was removed prior to power on
Install pan support
Err 8.3
Weight on pan support exceeds capacity
Remove weight from the pan support
Err 8.4
Pan support was removed during weighing
Re-install pan support
Err 8.5
Cannot perform temperature calibration
Cool down the analyzer
Err 8.6
Current temperature is higher than setting
Cool down the analyzer
Err 9.5
Factory calibration data corrupted
Contact an Ohaus-authorized dealer
Err 10.3
Temperature sensor out of range – high
Contact an Ohaus-authorized dealer
Err 10.4
Temperature sensor out of range – low
Contact an Ohaus-authorized dealer
Err 10.5
Temperature sensor defective
Contact an Ohaus-authorized dealer
Err 10.6
Temperature sensor defective
Contact an Ohaus-authorized dealer
Err 10.7
Temperature out of control
Contact an Ohaus-authorized dealer
Err 11.0
AC power is unstable
Use a power line conditioner
Err 53
EEPROM checksum error
Contact an Ohaus-authorized dealer
Err 54
Loading of calibration parameters failed
Contact an Ohaus-authorized dealer
Service Information
If the troubleshooting section does not resolve or describe your problem, contact your authorized OHAUS service agent.
Please visit our web site: www.ohaus.com to locate the OHAUS office nearest you.
Accessories
Description
Part No.
Anti Theft Device
80850043
Temp Calibration Kit
11113857
In Use Cover
80252475
Sample Pan Handler
80252476
Reusable Wire Cage
80252477
Reusable Sample Pan 7mm, 3pcs
80252478
Reusable Sample Pan 14mm, 3pcs
80252479
Aluminum Sample Pans, 80pcs
80850086
Pads, Glass Fiber, 200pcs
80850087
Reusable Sample Pan, 3pcs
80850088
Description
Part No.
Printer Impact, SF40A
30045641
Printer Impact, SF40A, AM
30064203
Printer Impact, SF40A, EU
30064202
Cable, RS9 M/F,1.5m
12122603
STP103 Thermal Printer, US
80251992
STP103 Thermal Printer, EU
80251993
STP103 Thermal Printer, UK
80251994
Cable, STP103
80252581
Содержание MB27
Страница 2: ......
Страница 14: ...EN 12 MB27...
Страница 26: ...ES 12 MB27...
Страница 38: ...FR 12 MB27...
Страница 50: ...DE 12 MB27...
Страница 62: ...IT 12 MB27...
Страница 63: ......