MOUNTING SUPPORT
• Keep attention through adjustment of the lid lock, that the lid lies straight and
continuous on the lid sealing. After adjustment tighten the screw well.
Adjustment:
1.) Lay the lid onto the centrifuges seal. Afterwards tighten the screws of the hinges
under a little pressure onto the lid.
2.) Put some washers underneath of the lid lock, in that way that the lid latch is
centric to the closing bar.
3.) Adjust the high of the lid lock, that the distance between the lid up to the housing
is about 8 mm to 9 mm.
4.) Bent the steel nib of the upper micro switch in that way, that it is given the
signal after the lid latch has pressed the steel nib under the level of the bar. The
motor of the lid lock should not be blocked during the closing of the lock.
Otherwise you have to repeat the steps 3.) and 4.).
5.) Perhaps you have to bent the steel nibs of the other micro switches of the end
stops into their position.
6.) Tighten all the screws well.
3.12 Replacing the rotor sensor
• Remove the lower covering of the front housing as described in chapter 3.1.
• Remove all electrical supply on the control board (1).
• Remove the motor cover.
• Remove both plastic screws from the rotor sensor (aerial) and remove the sensor.
While doing this put the connection cable through the intended bore.
1
1
Plastic screws