EN-13
FD Series
3.7 Scale Settings
Enter the Menu mode. Proceed to a menu item and use the
Yes
and
No
buttons to view and select
available scale settings. When the desired changes have been made, press
Exit
to return to weighing.
Refer to Section 3.4 for more information on how to enter and navigate the menus.
3.7.1
Calibration sub-menu
Note
: Calibration is not available when the scale has been set to Legal for Trade operation (LFT set to
ON in the Setup sub-menu, refer to Section 3.7.2).
Calibration Types – The FD Series’ calibration can be adjusted in two ways, Span and Linearity. Span
calibration uses two calibration points to adjust the sensitivity of the scale. Linearity calibration uses
three calibration points to correct for non-linear weighing results.
Calibration Weights – Before beginning calibration, make sure that the required calibration weights are
available. The model-specific calibration points are listed in the table below.
TABLE 3-3. CALIBRATION POINTS.
Notes
:
1. Default calibration points are shown in bold print.
2. Calibration Units kg or lb are selected in the Setup sub-menu (Section 3.7.2.)
3. ASTM Class 4 or OIML Class F2 weights are required for calibration.
Linearity Calibration Points
Model
Calibration Unit
Span Calibration Points
FD3/
FD3H
FD6/
FD6H
FD15/
FD15H
kg
lb
kg
lb
kg
lb
1, 2 or
3 kg
2, 5 or
6 lb
2, 5 or
6 kg
5, 10 or
15 lb
5, 10 or
15 kg
10, 20 or
30 lb
1
or 2 and
3 kg
2
or 4 and
6 lb
2
or 4 and
6 kg
5
or 10 and
15 lb
5
or 10 and
15 kg
10
or 20 and
30 lb