![OHAUS Explorer E1F110 Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/ohaus/explorer-e1f110/explorer-e1f110_datasheet_4266428002.webp)
Setting up the balance (continued):
• Getting into Print setup mode.
Press the “Setup” button.
Press the “right” button until “PRINT” is displayed then press the “Enter” button.
• Setting the balance to autoprint.
“AUTOPRT” should now be displayed, press the “Enter” button.
Press the up or down buttons until “CONT” is displayed then press the “Enter” button.
• Returning to normal operation.
Press the left or right buttons until “Exit” is displayed, press the “Enter” button.
The balance is now set up to work with the LogIT Serial Balance Adapter - See Connecting the Balance for details of the
lead required to connect the balance to the LogIT Adapter.
Connecting the balance:
A cable must be made to connect the balance serial port to the LogIT Serial Adapter using the wiring information below.
Explorer RS232 port, 9 way D Socket
Balance Adapter, 9 Way D Socket
pin
2
pin 2
pin 5
pin 6
pin 7
pin 5
Note:
These connection instructions are designed for use with the 9 pin boxed version of the LogIT Serial Balance
Adapter. If using the original 25 pin lead type connection you need to use adapter pin 3 instead of pin 2 (data out) and
adapter pin 7 instead of pin 5 (ground).
Serial Balance Adapter Data Sheet
OH-2
© DCP Microdevelopments Ltd
www.dcpmicro.com
12/12/00
Connect balance pin 5 to pin 6 to bypass handshaking