A-2
APPENDIX A EB SERVICE MODES/SERVICE CALIBRATION
A.2
SERVICE CALIBRATION (Cont.)
3. Add the displayed weight and wait until the display shows the third weight to add (full capacity).
4. After accepting the weight, the scale will restart the scale check process. Remove the weight
while the self-check is occurring.
5. The display will show 0.0. Turn off the scale.
6. Refer to Figure A-1 and return the jumper to the normal position.
A.3
SCALE CONFIGURATION
1. After performing paragraph A.1 and xx000 is displayed, press COUNT button to step through
the available settings for points of resolution as shown on table A-1, disregard any settings
not shown in the table.
TABLE A-1. SCALE RESOLUTION SETTINGS.
MODEL
RESOLUTION
CAPACITY
3
30000
3000.0
6
60000
6000.0
15
15000
15000
30
30000
30000
2. When the correct value is displayed, press the
TARE
button to accept.
3. The display will now show the capacity, press the
COUNT
button to step through the available
settings.
4. When the correct value is shown, press the
TARE
button to accept. The display will now show
0.0.
5. After the configuration of the scale has been done, it must be calibrated, perform procedure
in paragraph A.2.