![OHAUS EB SERIES Скачать руководство пользователя страница 17](http://html2.mh-extra.com/html/ohaus/eb-series/eb-series_service-manual_4243985017.webp)
CHAPTER 2 DIAGNOSIS
EB Series Scales Series Service Manual
2-5
Ohaus Corporation www.ohaus.com
2.4
EB SERIES SCALES ERROR CODE TABLE
The EB Series Scales is equipped with software which will display an error condition when it
occurs. Review the listed codes and follow instructions to correct the problem.
TABLE 2-7. EB SERIES SCALES ERROR CODES
ERROR CODE
PROBABLE CAUSE
REMEDY
E1, E2, E3
displayed
The pan is placed incorrectly.
Zero limit exceeded during
power up.
Load Cell Assembly Defective.
Main PC Board defective
Reposition the Pan correctly.
Ensure the Pan is empty before
switching the scale on.
Test the Load Cell (Section 2.7). If
defective replace it (Section 3.4).
After repair, proceed with Performance
Tests in Chapter 4.
Test the Main PC Board (Section 2.8).
If defective, replace it (Section 3.3).
After repair, proceed with Performance
Tests in Chapter 4.
--OL
—
displayed
Load exceeds scale capacity.
Load cell defective.
Main PCB is defective.
Reduce the load on the Pan.
Test the Load Cell (Section 2.7). If
defective replace it (Section 3.4).
After repair, proceed with Performance
Tests in Chapter 4.
Test the Main PC Board (Section 2.8).
If defective, replace it (Section 3.3).
After repair, proceed with Performance
Tests in Chapter 4.
Содержание EB SERIES
Страница 2: ......
Страница 4: ......
Страница 8: ...TABLE OF CONTENTS Ohaus Corporation www ohaus com iv EB Series Scales Service Manual...
Страница 24: ...CHAPTER 2 DIAGNOSIS Ohaus Corporation www ohaus com 2 12 EB Series Scales Series Service Manual...
Страница 48: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 4 EB Series Scales Service Manual...
Страница 52: ......
Страница 53: ...80251392 P N 80251392 D SERVICE MANUAL EB SERIES SCALES...