background image

 

Ohaus Corporation  www.ohaus.com 

Defender

™ Series Bases Service Manual 

 

1.8 

Specifications 

The technical data is valid under the following ambient conditions: 

Operating temperature: -10ºC to +40 ºC (14 °F to 104 °F) 

Relative humidity: 10% to 90% relative humidity, non-condensing 

Altitude: Up to 2,000m 

 

Note:

 Not all models may be available in your area. Check with your local OHAUS representative.

 

 

TABLE 1-1.  SPECIFICATIONS FOR DEFENDER SERIES BASES 

 

MODEL 

i-D3K1S 

i-D6K1S 

i-D15K1R 

i-D30K1R 

i-D60K1L 

i-D150K1L 

Capacity 

3 kg 

6 kg 

15 kg 

30 kg 

60 kg 

150 kg 

Approved Resolution 

OIML 3000e 

Safe Overload Capacity 

150% of capacity 

Pan Dimensions 

254 x 254 mm 

10 x 10 in 

305 x 305 mm 

12 x 12 in 

400 x 500 mm 

15.7 x 19.6 in 

Base Construction 

Stainless Steel platform with stainless steel frame and rubber leveling feet 

Repeatability (std.deviation) 

2d 

Linearity 

± 2d 

Load Cell Capacity 

10 kg 

10 kg 

20 kg 

50 kg 

100 kg 

200 kg 

Load Cell Cable 

1 m L x 6-wire 

2 m L x 6-wire 

Load Cell Type 

350 Ohm, SST, single point 

Load Cell Excitation 

3.3-15V DC/AC 

Load Cell Rated Output 

2mV/V 

Load Cell Protection 

IP69K 

Net Weight 

6 kg / 13.2 lb 

7.5 kg / 16.5 lb 

16 kg / 35.3 lb 

Shipping Weight 

7 kg / 15.4 lb 

8.5 kg / 18.7 lb 

18 kg / 39.7 lb 

 

 

 

Содержание Defender 6000 Series

Страница 1: ...ustin Ft Worth NicolScales com 800 225 8181 Contact Us Nicol Scales Measurement is an ISO Accredited Calibration Company that has provided calibration repair and sales of all types of weighing and mea...

Страница 2: ...DefenderTM 6000 Series Base Service Manual...

Страница 3: ...hecking Load Cells for Trouble 11 CHAPTER 3 MAINTENANCE AND REPAIR PROCEDURES 12 3 1 Preventive Maintenance 12 3 1 1 Preventive Maintenance Checklist 12 3 2 Replacing the Load Cell 12 3 3 Overload Sto...

Страница 4: ...uides for operational and performance tests and adjustments Chapter 5 Parts Identification Contains exploded views of Defender series Bases and parts identifying all serviceable components with parts...

Страница 5: ...intended for indoor use and should only be operated in dry locations Operate the equipment only under ambient conditions specified in the user instructions Do not operate the equipment in hazardous o...

Страница 6: ...irements of ASTM Class or OIML Mass values 3kg to 600kg depending on the Base model The full scale rated capacity of the Base should be used for Span calibration 1 7 Service Strategy The Load Cells us...

Страница 7: ...Capacity 3 kg 6 kg 15 kg 30 kg 60 kg 150 kg Approved Resolution OIML 3000e Safe Overload Capacity 150 of capacity Pan Dimensions 254 x 254 mm 10 x 10 in 305 x 305 mm 12 x 12 in 400 x 500 mm 15 7 x 19...

Страница 8: ...400 x 500 mm 15 7 x 19 6 in Base Construction Stainless Steel platform with stainless steel frame and rubber leveling feet Repeatability std deviation 2d Linearity 2d Load Cell Capacity 10 kg 20 kg 50...

Страница 9: ...Pan Dimensions 254x 254 mm 305 x 305 mm 400 x 500 mm 600 x 800 mm Base Construction Stainless Steel platform with stainless steel frame and rubber leveling feet Load Cell Capacity 10 kg 10 kg 20 kg 5...

Страница 10: ...luences such as moisture physical damage or electrical interference The strain gauges are wired into a Wheatstone Bridge Circuit See Figure 1 1 Figure 1 1 Downward force on the Platform bends the beam...

Страница 11: ...eight is loaded to the Load Cell For example a 500kg Load Cell with 2mV V output and 5V excitation would have a change of 0 02mV per kg change in the load that is 10mV 500 0 02mV per kg At a load of 2...

Страница 12: ...com 10 Defender Series Bases Service Manual Type2 For models i D150K1L i D300K1VZH i D600K1VZH bases FUNCTION WIRE COLOR Excitation Green Excitation Black Signal White Signal Red Sense Light Yellow Se...

Страница 13: ...unction Make sure that the proper Platform is supplied with the Base Replace all damaged parts See Chapter 5 for parts list Remove the Platform from the Base Check that the Overload Stops are not touc...

Страница 14: ...efore using the base 3 1 1 Preventive Maintenance Checklist The scale should be inspected and checked regularly as follows 1 Clean the base 2 Remove the Platform to inspect and clean the area beneath...

Страница 15: ...m Frame and the Load Cell Spacer aside 4 Remove the Top Load Cell Mounting Bolts washers and spacer The Load Cell assembly can now be removed from the Top Frame 5 Remove the Mounting Block Bolts from...

Страница 16: ...t and re check the Gap 4 Re adjust if necessary 5 Cover the Platform and check for full capacity See Table 3 1 for Gap settings TABLE 3 1 OVERLOAD GAP SETTINGS FOR DEFENDER SERIES BASES Base Model Bas...

Страница 17: ...ARTTENTION Make sure the test area is free from drafts and that the base rests on a level and vibration free surface 4 2 Load Cell Resistance Checks Use an ohm meter to measure across each pair of wi...

Страница 18: ...nnect the excitation signal shield and sensing wires if provided to the Indicator Calibrate the Indicator to read the load in the units required for example kilograms or grams Consult the Indicator in...

Страница 19: ...NG 3 Using an mV meter measure and record the signal voltage 4 Obtain the mV V output figure from the Load Cell data supplied with the Load Cell and compare the signal changes seen with the theoretica...

Страница 20: ...ssed the Load Cell must be adjusted or replaced Prior to starting this test the Base must be connected to a properly functioning Indicator with capacity and readability values set according to the spe...

Страница 21: ...Move the weight to point D and record the reading 7 If the reading at point A is negative file at points 1 and 4 AT AN ANGLE 8 If the reading at point B is negative file at points 1 and 2 STRAIGHT ACR...

Страница 22: ...ll cases where a part is replaced the base must be thoroughly checked after the replacement is made The base MUST meet the parameters of all applicable specifications in this manual If further technic...

Страница 23: ...R S Base Figure 5 1 Defender 6000 series Base R S Base Drawing Item Description Comments 1 Plate S Base i DT61 Plate R Base i DT61 2 Loadcell 1 PW15AHC3 10KG 1 i D3K1S i D5K1S i D6K1S i D3K1SZH i D6K1...

Страница 24: ...L Base Drawing Item Description Comments 1 Pan 400x500 SUS316 i DT61 2 Loadcell 1 PW15AHC3 100KG 1 i D50K1L i D60K1L i D60K1LZH Loadcell SSH 200kg 2 5m i D150K1L Loadcell SSH 300kg 2 5m i D150K1LZH 3...

Страница 25: ...ll SSH 500kg 2 5m i D300K1VZH Loadcell SSH 1t 2 5m i D600K1VZH 3 Level Kit D52 SST Base 4 Hardware Kit V Base i DT61 5 Foot 4 S R L base i DT61 6 Rubber Support SST Base i DT61 NS Box V D52 NS Box Com...

Страница 26: ...ugh typically there may be a small offset voltage at zero load Over the full rated capacity of the load cell the mV output will change The amount by which it changes depends upon the resistance change...

Страница 27: ...S Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 1 973 377 9000 Fax 1 973 944 7177 With offices worldwide www ohaus com 30655626 P N 30655626A 2021 OHAUS Corporation all rights reser...

Отзывы: