OHAUS C51XE6R Скачать руководство пользователя страница 13

CATAPULT 5000 SERIES SCALES                                                                                                          EN-7 

 

CAUTION  

RISK OF EXPLOSION CAN OCCUR IF REPLACED 
WITH THE WRONG TYPE OR CONNECTED 
IMPROPERLY.  

 
                                         

  Dispose of the lead acid battery according to local laws and regulations. 
 

 

2.2.4  RS232 interface Cable to Indicator (Optional) 

 

Connect the optional RS232 cable to the RS232 connector Figure 1-1, item 5.  

Note:

 For installation instructions, please refer to optional RS232 user manual. 

Pin 

Connection 

N/C 

 

TXD 

 

RXD 

 

N/C 

 

GND 

 

N/C 

 

N/C

 

N/C

 

N/C

 

                                                                                    Figure 2-1. RS232 Pins

 
 
 

2.3 

Internal Connections 

 

Some connections require the housing to be opened.  
 

2.3.1  Opening the Housing

 

 

CAUTION: ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO THE 
INDICATOR BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS. THE HOUSING 
SHOULD ONLY BE OPENED BY AUTHORIZED AND QUALIFIED PERSONNEL, SUCH AS AN 
ELECTRICAL TECHNICIAN. 

 
Remove the four Phillips head screws from the rear housing.  
Open the housing being careful not to disturb the internal connections.  
Once all connections are made, reattach the front housing. 
 

2.3.2  Jumper Connections  

For a 4-wire load cell with no sense wires: Jumpers W1 and W2 must be shorted.  
For a 6-wire load cell that includes sense wires, Jumpers W1 and W2 must be opened.  

Note

: 6-wire load cell setting is default. 

 

 

Figure 2-2. Jumper Connections. 

 

After wiring is completed and jumpers are in place, replace the indicator housing screws. 
 

Содержание C51XE6R

Страница 1: ...Catapult 5000 Series Scales Instruction Manual...

Страница 2: ......

Страница 3: ...nal 6 2 2 4 RS232 interface Cable to Indicator Optional 7 2 3 Internal Connections 7 2 3 1 Opening the Housing 7 2 3 2 Jumper Connections 7 2 4 Levelling the Instrument 8 2 5 Wall Base Mount 8 2 6 Ini...

Страница 4: ...set 19 3 7 2 Kilogram Unit 19 3 7 3 Gram Unit 19 3 7 4 Pound Unit 19 3 7 5 End Unit 19 3 8 Print Menu appears only with RS232 option installed 20 3 8 1 Reset 20 3 8 2 Baud 20 3 8 3 Parity 21 3 8 4 Sto...

Страница 5: ...ERIAL COMMUNICATION 25 5 1 Interface Commands 25 5 2 Output Format 25 5 3 Printout Examples 26 6 MAINTENANCE 27 6 1 Cleaning 27 6 2 Troubleshooting 27 6 3 Service Information 27 7 TECHNICAL DATA 27 7...

Страница 6: ......

Страница 7: ...product Note For useful information about the product Warning Symbols General Hazard Electrical Shock Hazard 1 2 Safety Precautions For safe and dependable operation of this equipment please comply wi...

Страница 8: ...view of Parts and Controls Item Description 1 Front Housing 2 Control Panel 3 Power Receptacle 4 Load Cell Cable Connector 5 RS232 Connector optional 6 Data Label 7 FCC Information 8 Rear Housing 9 Mo...

Страница 9: ...ls Cont Main PC Board Item Description 1 Power Connector J1 2 Battery Connector J9 3 Option Connector J2 4 LFT Switch 5 4 6 Lines Sense Jumper W1 located on the other side of PCB 6 Load Cell Connector...

Страница 10: ...t Item Description 1 Pound Kilogram Gram symbols 2 Battery function symbol 3 TARE Menu button 4 FUNCTION Mode button 5 Accumulation Symbol 6 NET function symbol 7 PRINT Units button 8 Dynamic symbol 9...

Страница 11: ...ress Off Turns the Indicator on or off Units Changes the weighing Unit Mode Allows changing the application mode Press and hold allows scrolling through modes Menu Enter the User menu Menu Function Sh...

Страница 12: ...or the power cord is removed Attention Before using the indicator for the first time the internal rechargeable battery should be fully charged for up to 12 hours The indicator can be operated during...

Страница 13: ...ing the Housing CAUTION ELECTRICAL SHOCK HAZARD REMOVE ALL POWER CONNECTIONS TO THE INDICATOR BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS THE HOUSING SHOULD ONLY BE OPENED BY AUTHORIZED AND QUALIF...

Страница 14: ...l each time its location is changed 2 5 Wall Base Mount A Wall Base mounting bracket and screws are provided Base mounting Models C51XE6R C51XE30R C51XE50R C51XE200R Models C51XE30L C51XE50L C51XE100L...

Страница 15: ...20 2d DYNAMIC GRAM 7 EVEN POWER ON UNIT 5d OFF OFF 7 ODD AUTO FILTER MANUAL ON 7 NONE GRAM LOW SEMI MAN LB 8 NONE KILOGRAM MED AUTO OFF STOP ZERO RANGE HI END MODE ON 1 0 AZT END UNIT 2 2 OFF HANDSHA...

Страница 16: ...set to On 4 Allow the Indicator to warm up for approximately 5 minutes after stabilizing to room temperature 5 To abort calibration press the Exit button anytime during the calibration process 6 Make...

Страница 17: ...ration Linearity calibration uses 3 calibration points The first calibration point is established with no weight on the scale The second calibration point is established at approximately half capacity...

Страница 18: ...s in gravity Attention Changing the GEO Factor alters the calibration The GEO value was set at the factory and should only be changed by an authorized manufacturer s representative or certified verifi...

Страница 19: ...14 13 13 12 12 11 11 10 40 05 41 12 16 15 15 14 14 13 13 12 12 11 11 41 12 42 19 16 16 15 15 14 14 13 13 12 12 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15 15 14 14 13 1...

Страница 20: ...ty Graduation Zero Range and settings are not reset 3 4 2 Capacity Set the scale capacity from 5 to 20000 Refer to the Setup Table 3 3 for available settings CAUTION Once the Capacity value has been c...

Страница 21: ...CATAPULT 5000 SERIES SCALES EN 15 TABLE 3 3 SETUP AND CALIBRATION VALUES 6 15 0 0005 0 001 0 002 0 005 6...

Страница 22: ...on Auto last unit in use when power was turned off kg g lb 3 4 5 Zero Range Set the percentage of scale capacity that may be zeroed 2 zero up to 2 percent of capacity 100 zero up to full capacity 3 4...

Страница 23: ...drift of 1 division per second has been exceeded 3 d the display will maintain zero until a drift of 3 divisions per second has been exceeded NOTE When the LFT menu item is set to ON the selections a...

Страница 24: ...resetting is initiated by pressing the FUNCTION key AUtO averaging is automatically initiated when the display is stable resetting is automatically initiated when the weight on the display is 1 divis...

Страница 25: ...1 Reset Set the Unit menu to the factory defaults NO not reset YES reset 3 7 2 Kilogram Unit Set the status OFF Disabled ON Enabled 3 7 3 Gram Unit Set the status OFF Disabled ON Enabled 3 7 4 Pound U...

Страница 26: ...nt 3 8 2 Baud Set the Baud rate 300 300 bps 600 600 bps 1200 1200 bps 2400 2400 bps 4800 4800 bps 9600 9600 bps 19200 19200 bps Reset No Yes Baud Rate 300 600 1200 2400 4800 9600 19200 Parity 7 Even 7...

Страница 27: ...w control method NONE no handshaking ON OFF XON XOFF software handshaking 3 8 6 Stable Only Set the print criteria OFF values are printed immediately ON values are only printed when the stability crit...

Страница 28: ...et weight is printed TARE OFF Tare weight is not printed ON Tare weight is printed UNIT OFF Unit is not printed ON Unit weight is printed MODE OFF Mode is not printed ON Mode is printed INFO OFF Info...

Страница 29: ...peration 4 3 Manual Tare When weighing an item that must be held in a container taring stores the container weight in memory Place the empty container on the scale example 0 5 kg and press the TARE bu...

Страница 30: ...accumulated data with no weight on pan press Function Mode key once The display blinks while showing the number of samples total weight Pressing the ZERO key zeros the display if required without aff...

Страница 31: ...0T clears tare if allowed PU Print current unit g kg lb PCS xU Set scale to unit x 1 g 2 kg 3 lb xM Set scale to mode x 1M weigh 2M count 3M total 4M dynamic M will scroll to next enabled mode PV V Ve...

Страница 32: ...Gross value line If Print Content Gross is ON and a tare value is entered 9 72 KG N Net value line If Print Content Net is ON and a tare value is entered 2 62 KG T Tare value line If Print Content Ta...

Страница 33: ...y icon flashing Low Battery error Battery is empty CAL E Calibration Error Calibration value outside allowable limits Lo rEF Low reference weight warning message Average Piece Weight too small Warning...

Страница 34: ...tainless steel Net Weight 4 kg 13 5 kg Shipping Weight 5 kg 15 kg Shipping Dimensions 400 x 365 x 210 mm 600 x 480 x 220 mm Scale Models C51XE100L C51XE200L C51XE30X C51XE50X C51XE100X C51XE200X Capac...

Страница 35: ...T NUMBER RS232 Kit 30101019 In Use Cover 30101017 SF40A Impact Printer Contact OHAUS Lead Acid battery 30251812 7 3 Drawings and Dimensions Figure 8 1 C51 Overall Dimensions Dimention Chart mm A B C D...

Страница 36: ...nt in EU Member States To comply with this Directive this device has been designed for safe removal of the batteries at end of life by a waste treatment facility Please dispose of this product in acco...

Страница 37: ...t has been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other...

Страница 38: ......

Страница 39: ......

Страница 40: ...ec des bureaux dans le monde entier Mit Niederlassungen weltweit Con uffici in tutto il mondo www ohaus com 30380009 P N 30380009B 2017 OHAUS Corporation all rights reserved todos los derechos reserva...