Switch On/Off
On
Off
Switch EC/TDS/SAL/RES Mode
EC
TDS
SAL
RES
EC Reference Temperature Setup
EC Temperature Coefficient Setup
Touch the value on the screen
to confirm your setting.
EC TDS Factor Setup
Touch the value on the screen
to confirm your setting.
EC Cell K Setup
Touch the value on the screen
to confirm your setting.
Note
: Please set the Cell K value first before EC calibration.
Calibration
Touch the
Calibration
icon on the
screen.
Calibrating
Press the
button when you
finish to continue.
Touch
Finish
to complete
calibration.
The calibration result will be
shown when you finish.
Note
: Predefined standard solution: 10 μS/cm, 84 μS/cm, 146.5 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm. Please set the Standard Solution first before EC calibration.
EC Standard Solution Setup
Temperature Unit Setup
Temperature Setup
Touch the value on the screen
to confirm your setting.
Measurement
Insert the probe into a sample.
Measuring
Completed
Touch the mode area on the
screen to select measurement
mode.
Store/Recall Data
Print:
Touch
the
Printer
icon on
the screen to print.
Recall measurement data:
Touch
the
Beaker
icon on the
screen to recall.
Note
: measurement data will
be automatically stored after
each operation.
a-AB33EC Quick Start Guide