CHAPTER 2 TROUBLESHOOTING
Symptom Possible
Cause(s) Remedy
Error 9.5
Calibration data not present.
Calibrate scale.
Replace PCB.
Error 53
EEPROM data incorrect.
Cycle power on and off.
If error continues, replace PCB.
CAL E
Calibration Error. Calibration
value outside
allowable limits.
Repeat calibration using correct
calibration
weights.
LOW rEF
In Pieces mode, the average
piece weight of the parts is
small (warning).
You may continue to weigh, but results
may not be accurate, because average
piece weight is smaller than can be
accurately counted.