OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
7
Setup
Viscometer
1. Place the unit on a solid, level surface away from any large motors or
strong magnetic fields.
2. The HTHP Viscometer is designed to test samples at elevated
temperatures and pressures. It requires an air supply of 100 PSI (150 PSI
maximum) to drive the pump and a water source (standard tap water) to
cool the unit and the test sample. The panel on the left-hand side of the
unit has 3 fittings, ¼" (6.35 mm) NPT, for the air supply, water source, and
drain.
The viscometer requires a constant air supply throughout the test.
3. Connect the DB-9 cable to the viscometer and the computer COM port 1.
This connection can be found on the right panel of the unit.
4. Connect the 68-pin cable to the viscometer and the computer. On the unit,
this connection can be found next to the DB-9 connection. There will be a
matching connection on the back panel of the computer.
5.
Make sure the viscometer power switch is in the “off” position. Connect
the power cord to the unit and an AC power source.
6. Turn the unit on. The power switch is located on the lower right panel.
Once the unit is turned on, you will immediately hear both fans operating
inside the tower and a beep from the electronics.
Before running a test on the viscometer, you must zero the compass
and then run a calibration procedure. For detailed instructions, see
page 18 for calibrating the unit and page 47 for zeroing the
compass.
Transducer Housing
Tower
Pressure Release
Valve
Test Cell
Data Acquisition
Cable
Serial Communication
Cable
Viewing Ports (Oil Level)