OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
14
Export Standard
Manual Calibration
On the DigiBASE Manual Control screen you can run a manual calibration
on a single standard. Refer to page 8 for more information on the Manual
Control screen.
1. Select “DigiBASE Manual Control” from the “Options” menu.
2. From the drop-down menu, select the standard you want to calibrate
(Potassium, Uranium, Thorium or Gamma API).
3. Enter a value for True Time and click the “Set True Time” button. The
standard calibration time is 10,000 seconds (2 hours, 46 minutes, 40
seconds) per standard.
4. Place the standard under the shield.
5. Click the “Start” button to start logging.
6. When the True Time has elapsed, logging will stop automatically. Click the
“Export Standard” button to save the calibration.