16. Place the cell in the heating jacket. Rotate the cell so that the pin in
the bottom of the heating well seats into the hole in the bottom of the
cell. This will anchor the cell inside the well and prevent the cell from
rotating as the valve stems are opened and closed. Once anchored, you
may rotate the cap, mag drive, or stem to better orient the valve and/or
thermocouple.
17. Place the cell temperature thermocouple into the thermometer well in the
cell.
The reading from the thermocouple will display on the “Cell Temperature”
Eurotherm. Allow the cell to heat for one hour. Start a 60 minute timer
now.
Control
Thermocouple
Bleeder Valve
Manifold Block
Valve Stem
Safety Pin
18. Plug the magnetic drive into the port on the back of the control box.
Temperature
Controller
Cell Temperature
Thermocouple
Pressure
Mag Drive
Quick-Connect
Fitting