Maintenance
1. One of the most important elements of trouble free operation is keeping
the interior of the test cell as clean as possible. Never insert a cement
covered cement mold into the test cell and periodically examine the test
cell to ensure that it is clean. High-pressure valves wear quickly when
exposed to fluids containing cement and other particulate matter. Also
make sure the air used for the air supply is clear.
2.
OFITE uses a high pressure filter to protect valve stems. It is
recommended that filtered water be used in the curing chamber to
prevent particulate matter from entering the pump and possibly causing
damage. Clean or replace these filters when fluid flow is reduced
from that of a clean system. Remember: Filters are inexpensive when
compared to the costs of replacing the components they were designed
to protect.
3. The test cell cap threads have been lubricated prior to shipment and
periodically should be re-lubricated.