
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
55
Calibration
Temperature
The thermocouple will require calibration quarterly or if it is replaced.
1.
Place the thermocouple in a dry block calibrator.
2. Start the calibration procedure:
a. For onboard control, touch Calibrate and then Temperature.
b. For software control, select Temperature from the Calibrate menu
3. Follow the instructions on the screen.