4-4. Output Voltage/Current Setting & Operating Command
This section describes output voltage and current limit control protocol based on remote
interface
VOLT <voltage>
Output voltage setting command.
> voltage input voltage value
ex) volt 10
set the 10V
VOLT?
Confirm present setting voltage.
Return value "voltage"
ex) volt?
return value "30.0"
CURR <current>
Output current setting command
> current input current value
ex) curr 1.5
set the 1.5V
CURR?
Confirm present setting voltage
Return value "current"
ex) curr?
return value "3.15"
OVP <0~Max Voltage>
You can set the OVP(Over voltage protection) Trip Level
> <0~Max Voltage> Input the Voltage value.
ex) ovp 5.2
Set the OVP Level to 5.2V
4-5. Protection Command (Option)
OVP?
You can check the OVP(Over Voltage Protection) Trip Level.
Return value "voltage"
ex) volt:prot?
return value "5.20"
OVP:TRIP?
You can check the occurrence of OVP Trip.
Return value "0" - Normal Output State
"1" - Output is blocked because of OVP Trip
OVP:CLE
You can clear the OVP Trip.
- 37 -