15 Cleaning, Di
s
infection and Maintenance
Instruction Manual Corvis® ST (G/72100/XXXX/EN 1019 Rev01)
57 / 80
Clean lens protection glass (3)
The openings in the housing for the optics are covered by protective glass
covers which must be kept dust and dirt-free.
Clean the lens protection glass with a lint-free cloth moistened with
alcohol.
Clean the air nozzle (2)
Attention
Risk of infection after examining a sick patient
If you perform a measurement on a sick patient, the air nozzle can be
contaminated.
Moisten a cotton swab with rubbing alcohol.
Clean the nozzle with the cotton swab.
Disinfect the device as needed, see
Check the glass part of the air nozzle from a diagonal angle for dust,
dirt, etc.
Blow off dust, foreign particles, etc. with a lens blower.
Afterwards, carefully wipe the glass with a cotton swab which you
have previously moistened with methanol or pure alcohol.
Note
Damage due to improper cleaning
Carefully wipe the air nozzle and do not rub the surface.
Make sure there are no foreign particles on the air nozzle when
wiping it. Otherwise, the glass could get scratched.
Check the glass afterwards.
Cleaning the housing
Wipe the head rest after each examination, and the housing as required.
Unplug the power cord. When disconnecting electrical connections,
pull on the respective plug and not on the cable itself.
When cleaning, use a damp cloth and make sure that no liquid enters
the Corvis® ST.
Clean the lens in front of the camera using a dry, lint-free cloth.
Clean the plastic surfaces and painted surfaces with the appropriate
cleaning agents.