
15 Cleaning, Di
s
infection and Maintenance
58 / 80
Instruction Manual Corvis® ST (G/72100/XXXX/EN 1019 Rev01)
Cleaning the display
Clean the display using a dry, lint-free cloth.
15.2 Di
s
infection
Required materials:
Disinfection and Cleaning Kit (included),
Alternatively: Pursept®-A Xpress disinfectant wipes,
Merz+Co company; D-60318 Frankfurt:
Tel: +49 69 1503 1; Fax: +49 (69) 596 21-50
E-mail: [email protected]
Disinfect the head rest and, if neccessary, the chin rest after each
examination.
Attention
Risk of infection after examining a sick patient
If you perform a measurement on a sick patient, the head rest and the
front cover can be contaminated.
Disinfect the front cover (2) and the housing as required.
1
Head rest
2
Front cover
Abb. 15-2: Disinfect
1
2