surge
GB / T 17626.5
± 1 kV
differential
mode voltage
± 2 kV common
mode voltage
Not applicable
Not applicable
Temporary voltage
on power input line
Drops, short
interruptions and
voltage changes
Turn into
GB / T 17626.11
<5% UT for 0.5
cycles
(On UT,> 95% dips)
40% UT for 5 cycles
(60% sag on UT)
70% UT for 25 cycles
(30% sag on UT)
<5% UT for 5s
(On UT,> 95% dips)
Not applicable
Not applicable
Power frequency
magnetic field (50
/ 60Hz)
GB / T 17626.8
3A/m
3A/m,50/60Hz
Power frequency
magnetic field should
have power frequency
magnetic field level
characteristics in a
typical place in a
typical commercial or
hospital environment
Note: UT refers to the AC network voltage before the test voltage is applied
Guidance and manufacturer's declaration-electromagnetic immunity
This product is intended to be used in the electromagnetic environment specified
below. The purchaser or user of this product should ensure that it is in this
electromagnetic environment.
use:
Immunity test
IEC 60601 test level
Coincidence
level
Electromagnetic
Environment-Guide
Radio frequency
conduction
GB / T 17626.6
3 V(Effective value)
150 kHz~80 MHz
Not applicable
Portable and mobile RF
communications
equipment should not be
used closer to any part
of this product,
including cables, than
the recommended
isolation distance.
This distance should be
calculated by a formula
corresponding to the