Quick Start
–
Bench Setup
It is important to test
icListen
when it is first unpacked and before each deployment, for best results.
The hydrophone works in air well enough to provide a good functional test. Arrange a work surface large
enough to hold the hydrophone, cables and the PC used to run the
Lucy
software.
3.1
Checklist
1.
Hydrophone to be tested (note its serial number).
2.
A PC (ideally the same portable one used to deploy the hydrophone), with the
Lucy
software
already installed. Ensure the PC has a suitable data port interface for the
icListen
instrument.
3.2
Steps
1.
Connect the
icListen
to the link cable, and the data link to the PC.
2.
Start up the
Lucy
program on the PC.
3.
Click the icListen Setup button in the lower right. In the popup window, select the Link tab and
chose the appropriate connection type. To find your
icListen
, click “Find All Units”
. To connect,
double click the unit from the found units list
.