C: Analog Output
013-20000-009-02-201510
49
2. Select an appropriate data source from the dropdown box located to the right of the Current
Output Source setting. For more info on valid sources, see
3. Select a valid lower bound for your represented data range by typing a number into the box
located to the right of the “Analog Current Lower Bound” setting. In some versions of the Viewer
software, this setting may be called “Current Limit Minimum”. For more information on selecting
a lower bound for your voltage, see
4. Select a valid upper bound for your represented data range by typing a number into the box
located to the right of the “Analog Current Upper Bound” setting. In some versions of the Viewer
software, this setting may be called “Current Limit Maximum”. For more information on
selecting a upper bound for your voltage, see
You can save your settings by enabling the Flash Write setting momentarily.
Selecting a Data Source
There are six potential data sources that can be driven by the NeoFox. The pressure mode is not currently
supported as an analog data source.