Appendix A
Specifications
Specification
Value
Parameters Measured
Gasseous oxygen partial pressure, dissolved oxygen concentration,
temperature, pressure (via onboard pressure transducer)
Control Software
Windows 2000/XP/Vista (32 and 64-bit) and Windows 7 (32-bit) control
software with data logging capability; controls include: modulation
frequency, data rate, signal averaging, APD gain, analog gain, LED
intensity
Sampling Rate
Acquires data samples every 0.1 second
NOTE: This is not the response speed of the probe, just the rate at which
the probe is digitally sampled
Thermistor Probes
Closed-end stainless steel tube with thermistor sensor mounted in tip; liquid
immersible rugged design; 1/8” NPT fitting; temperature range 0 to 75 °C
Pressure Measurement
On-board pressure transducer monitors atmospheric pressure
Input Power
Approximately 2.5W @ 5V
Communications
USB, 0-5V analog output; 4-20mA analog output. RS232 is available on the
NeoFox GT model with an F or greater in the third digit of its serial number.
NeoFox Sport
Upload Interval
The NeoFox Sport uploads data from the NeoFox module at least once per
second in the Sport configuration. Alternate configurations may have lower
data rates.
Battery Life
In the NeoFox Sport configuration, a fully charged, gently used battery will
last for approximately three hours.
Содержание NeoFox
Страница 22: ...2 Installation 14 013 20000 009 02 201510 ...
Страница 42: ...4 Using NeoFox Sport 34 013 20000 009 02 201510 NeoFox Sport Menu Menu Tree Description ...
Страница 48: ...4 Using NeoFox Sport 40 013 20000 009 02 201510 ...
Страница 50: ...A Specifications 42 013 20000 009 02 201510 ...
Страница 54: ...B Maintenance 46 013 20000 009 02 201510 ...
Страница 63: ...C Analog Output 013 20000 009 02 201510 55 ...
Страница 64: ...C Analog Output 56 013 20000 009 02 201510 ...
Страница 66: ...D Calculating Drift Per Sensor 58 013 20000 009 02 201510 ...