Heat is not
functioning
2 & 3 blinking
lights
Integrated heat level indicator
DIRECTIONS FOR USE
OPTIONAL INTEGRATED HEATED TOP
INTEGRATED HEATED TOP TROUBLESHOOTING
Heating element locations
Possible causes-
• Plug is not properly plugged in.
• Wall outlet isn’t working properly.
• Damaged wires & terminals.
Solutions-
• Ensure that the all power cords are properly plugged into a
standard 120V or 230V wall outlet.
• Ensure heating element connectors (2) are properly connected to the control box.
• Check light indicators on the front of the unit to ensure power is supplied. If the indicator is not
illuminated after pressing it once try plugging the table into another outlet.
Possible causes-
• Two & three blinking lights on the control switch indicates that the heating element connector(s) are not
plugged in properly and/or damage to the terminals and/or connectors has occured.
Solutions-
• Disconnect power supply to the control box (located on the underside of the table top) and wall plug.
Disconnect all heating element connectors from the control box. Reconnect the heating element
connectors first then plug the power cord into the wall outlet.
• If problems continue contact Customer Service at 717-235-6807.
USING THE INTEGRATED HEATED TOP:
The table comes with the option for a fully integrated heated top pad. Simply press the heating indicator
located on the right side rail of the table to the desired heat level.