DIRECTIONS FOR USE
INTEGRATED HEATED TOP TROUBLESHOOTING
Heat is not
functioning
2 & 3
blinking
lights
Possible causes-
• Plug is not properly plugged in.
• Wall outlet isn’t working properly.
• Damaged wires & terminals.
Solutions-
• Ensure that the all power cords are properly plugged into a
standard 120V or 230V wall outlet.
• Ensure heating element connectors (2) are properly connected to the controller.
• Check light indicators on the front of the unit to ensure power is supplied. If the indicator is not
illuminated after pressing it once try plugging the table into another outlet.
Possible causes-
• Two & three blinking lights on the control switch indicates that the heating element connector(s) are not
plugged in properly and/or damage to the terminals and/or connectors has occured.
Solutions-
• Disconnect power supply to the control box (located on the underside of the table top) and wall
plug. Disconnect all heating element connectors from the controller. Reconnect the heating element
connectors first then plug the power cord into the wall outlet.
• If problems continue contact
Customer Service at 717-235-6807
.