DIRECTIONS FOR USE

INTEGRATED HEATED TOP TROUBLESHOOTING

Heat is not 

functioning

 

2 & 3 

blinking 

lights

 

Possible causes-  

•  Plug is not properly plugged in.
•  Wall outlet isn’t working properly.
•  Damaged wires & terminals.

Solutions-

•  Ensure that the all power cords are properly plugged into a 

standard 120V or 230V wall outlet.

•  Ensure heating element connectors (2) are properly connected to the controller.

•  Check light indicators on the front of the unit to ensure power is supplied. If the indicator is not 

illuminated after pressing it once try plugging the table into another outlet.

Possible causes- 

•  Two & three blinking lights on the control switch indicates that the heating element connector(s) are not 

plugged in properly and/or damage to the terminals and/or connectors has occured.

Solutions-

•  Disconnect power supply to the control box (located on the underside of the table top) and wall 

plug. Disconnect all heating element connectors from the controller. Reconnect the heating element 

connectors first then plug the power cord into the wall outlet.

•  If problems continue contact 

Customer Service at 717-235-6807

.

19

Содержание Clinician Premiere Series

Страница 1: ...UAL www OAKWORKS com 717 235 6807 made in the USA with US imported parts 4 Section Electric Salon Top Shown with optional Adjustable Breast Comfort System 1 Section Flat Top Shown with optional Adjust...

Страница 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifica tions and the like documen...

Страница 3: ...ontrol Operations Flat Top Tables 9 Electric Salon Tables 10 Adjustable Breast Comfort ABC Systems Only 11 Hand Control Operations Flat Top Tables 12 Electric Salon Tables 13 Adjustable Breast Comfort...

Страница 4: ...ries transducers and cables other than those specified by the manufacturer may result in increased emissions or decreased immunity of the table PRODUCT USE DESCRIPTION IMPORTANT SAFETY INSTRUCTIONS 1...

Страница 5: ...g administered Do not lift table by the table top This can damage the table When lowering the table make sure there is nothing underneath the table top that can impede motion like stools cabinets acce...

Страница 6: ...ns Hand Control All powered motions Table Voltage Options 120V 60Hz 230V 50Hz Table Capacity 500 lbs 227 kg client and accessories Widths 29 74 cm 31 79 cm or 33 84 cm Length 73 185 cm Padding 4 10 cm...

Страница 7: ...ptions 120V 60Hz 230V 50Hz Table Capacity 500 lbs 227 kg client and accessories Widths 29 74 cm 31 79 cm or 33 84 cm Length 73 185 cm Padding 4 10 cm or 6 15 cm Aerocel Foam Table Weight 170 lbs 77 kg...

Страница 8: ...he controls again If one or more of the controls still do not work contact Customer Service at 717 235 6807 to arrange replacement IF THE TABLE IS MAKING NOISE DURING HEIGHT ADJUSTMENT CAUTION Disconn...

Страница 9: ...s equipped with a cord having an equipment grounding conductor and a grounding pin The pin must be plugged into an appropri ate outlet that is properly installed and grounded in accordance with all lo...

Страница 10: ...movable for accessories Head end not removable if the table has integrated heat This product is equipped with an interior 120 volt 60Hz six plug accessory outlet The total electrical load of equipment...

Страница 11: ...sed to prevent unauthorized use Remove the key when not in use and store out of reach of children HYDRAULIC DAILY CYCLING The Clinician Premiere uses hydraulics and hydraulic fluids to adjust the tabl...

Страница 12: ...at Top tables offer height positioning with electronic ease Operate the foot control as shown below to raise or lower the height of the table DIRECTIONS FOR USE FOOT CONTROL OPERATIONS FLAT TOP TABLES...

Страница 13: ...for holding anything Damage can occur to the table The Oakworks Clinician Premiere Electric Salon Tables offer height backrest and footrest positioning with electronic ease Operate the foot control a...

Страница 14: ...stem and your client The Oakworks Adjustable Breast Comfort ABC System is designed to lower the pressure on breasts and pectoral muscles allowing the client to relax and create a more comfortable posi...

Страница 15: ...positioning with electronic ease Operate the hand control as shown below to raise or lower the height of the table DIRECTIONS FOR USE HAND CONTROL OPERATIONS FLAT TOP TABLES P N 71567 1 P N 71568 1 C...

Страница 16: ...g anything Damage can occur to the table The Oakworks Clinician Premiere Electric Salon Tables offer height backrest and footrest positioning with electronic ease Operate the hand control as shown bel...

Страница 17: ...stem is designed to lower the pressure on breasts and pectoral muscles allowing the client to relax and create a more comfortable position for their spine The height of the cushion is easily adjusted...

Страница 18: ...bs 86 kg Four people are required to move the table safely Do not use the table to transport a client DIRECTIONS FOR USE Position at low height TABLE LEVELING INSTRUCTIONS 1 Place the table in the loc...

Страница 19: ...t level before they are on the table NOTE Breast size is not the only determinate of cushion location Every person s comfort spot is different regardless of breast size HEAD REST USE IN SUPINE If you...

Страница 20: ...the QuickLock Face Rest gently pull the headrest by the dowels until it comes free Set it safely aside or store it inside the under table cabinet until needed DIRECTIONS FOR USE QuickLockTM FACE REST...

Страница 21: ...lly integrated heated top pad Simply press the heating indicator located at the head end of the table to the desired heat level The controller for the Integrated Heated Top is located underneath the s...

Страница 22: ...ower is supplied If the indicator is not illuminated after pressing it once try plugging the table into another outlet Possible causes Two three blinking lights on the control switch indicates that th...

Страница 23: ...d wipe for cleaners disinfectants to ensure best distribution of disinfectant for the required kill time without leaving excess residue and or overexposing components therefore minimizing the potentia...

Страница 24: ...wet for the disinfectant manufacturer s recommended contact time 4 Throw the used wipe s away and allow the table to dry INSPECTIONS RECOMMENDED REGULAR INSPECTIONS monthly or local standard Check fo...

Страница 25: ...ted on the underside of the foot section of the table top Model and Serial numbers always start with a letter 16796025 CPSA4EF1T411USXXPW Ser PAD123456 Clinician Premiere 4 Sect SA 4EF 1 T 4 11 US XX...

Страница 26: ...otection Class 1 Equipment Class 1 Equipment Tabletop Applied Part Type B Applied Part Type B Applied Part Ingress Protection Rating IPX0 IPX0 Made of Operation Intermittent Operation MAX 2 minutes ON...

Страница 27: ...THIS PAGE IS INTENTIONALLY LEFT BLANK 24...

Страница 28: ...6807 FAX 717 235 6798 www oakworks com USER MANUAL 3034177 CONFORMS TO UL STD 962 Certified to CAN CSA STD C22 2 68 Original Instructions Manual Part Number SPA RA PM 141 Revision level 2 Revision Dat...

Отзывы: