Instruction Manual 

 

Temp-360  

RTD Data logging Thermometer 

 

 

 

 

 

 

 

 

 

68X526601 Rev 2 06/10

Part of Thermo Fisher Scientific 

Содержание Temp-360

Страница 1: ...Instruction Manual Temp 360 RTD Data logging Thermometer 68X526601 Rev 2 06 10 Part of Thermo Fisher Scientific...

Страница 2: ......

Страница 3: ...DISPLAY OVERVIEW 11 10 MEASUREMENT MODE 13 11 HOLD FUNCTON 14 12 MIN MAX AND AVE FUNCTION 14 13 DATA LOGGING 14 14 SETUP MODE 15 15 GENERAL SETUP SCREEN 16 16 CALIBRATION SCREEN 18 17 ALARMS SCREEN 20...

Страница 4: ......

Страница 5: ...logging for up to 2000 points USB output Operator selection of Celsius or Fahrenheit scale Resolution of up to 0 01 C F Large backlit LCD and dot matrix graphic LCD display Hold feature for temporari...

Страница 6: ...zardous atmosphere batteries must only be changed in an area known to be non hazardous CAUTION 1 Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas 2 Weak batte...

Страница 7: ...01 C F From 99 99 to 99 99 C F 0 1 C F From 100 0 to 999 9 C F 1 C F Above 1000 C F Accuracy 0 1 C 0 2 F From 330 0 to 100 C F 0 03 C 0 06 F From 99 99 to 99 99 C F 0 1 C 0 2 F From 100 0 to 999 9 C F...

Страница 8: ...pin DIN connector Input Protection 50V rms Storage 40 C to 65 C 40 F to 149 F Humidity 10 to 90 non condensing Battery Life Size Three AA 1 5V Alkaline Life 400 hours continuous typical without backli...

Страница 9: ...5 Ingress Protection Meets IEC 529 IP 54 for dust and water resistant enclosures with probe attached CE Compliance EN61326 1 A1 1998 EU EMC Directive...

Страница 10: ...ty bars flashing More than 2 85 V Adaptor power supply Main power supply connected USB power supply USB power Selected settings are stored in memory and will remain in memory even after power is turne...

Страница 11: ...armor slide in from the top of meter before pushing the bottom edges of meter down to set it into position Lift up the stand at the back of meter for bench top applications if necessary 2 To remove th...

Страница 12: ...6 ASSEMBLING OPTIONAL HANDSFREE ACCESSORIES You can use the optional magnets and strap in the Handsfree Kit accessories for hands free operations 8...

Страница 13: ...gs Insert the 3 pin plug into the mating connector on the top of the instrument If no probe is connected the display will read open RTDs are sensitive at the tip or sensing element When taking measure...

Страница 14: ...e display on off light Turns meter on and off press and hold for 3 seconds to turn off Press momentarily to turn on backlight recall Recalls and steps through stored readings log Stores current measur...

Страница 15: ...9 DISPLAY OVERVIEW The dot matrix display features a large primary display smaller secondary displays for channel info or min max ave and helpful annuciators for added measurement data 11...

Страница 16: ...cator 8 Data logging indicator 9 PC data logging indicator 10 Main reading display 11 Current reading unit indicator 12 Min Max Avg elapsed time 13 Current Min Max Avg reading indicator 14 Meter loggi...

Страница 17: ...l display the measured value in the primary display Pressing the F2 key will toggle reading between F and C display Pressing the F1 key initiates and toggles through the Minimum Maximum and Average Re...

Страница 18: ...ge that occurs The minimum and maximum values are sensed and automatically stored To exit and clear this function press the F3 key to access the Menu functions See the CLEAR RESET SCREEN section for m...

Страница 19: ...ey pad to scroll through options To enter a setup screen press the Select key F1 To return to the measurement mode press the Meas key F3 The following menu options are listed 1 General setup 2 User fi...

Страница 20: ...Press the F1 key to indicate that you want to change the setting of the current parameter or press the recall key or log key to move to the next parameter Press the recall key or log key to change th...

Страница 21: ...The screen below is used to reset change the password If the user forgets the password 5586 can be used to reset to a new value On the second screen you can set the time and date 17...

Страница 22: ...emperature point to compensate for RTD offset error It is NOT necessary to perform a field calibration to obtain the specified meter accuracy Use the field calibration feature to improve thermometer p...

Страница 23: ...he offset range of 10 00 C or 18 00 F Slope adjusts at two points The two calibration points must be at least 40 00 C 72 00 F apart The second calibration point should be at a higher value than the fi...

Страница 24: ...tings available in the alarm setting options MEASUREMENT ALARM Disable or enable the alarm by pressing the recall or log key and the F1 key to accept Increase or decrease the individual limit by press...

Страница 25: ...COUNTDOWN ALARM You can enable disable the countdown alarm and set the countdown time from 5 seconds to 1 hour After setting enabling the countdown alarm the measurement screen will look like this 21...

Страница 26: ...art the countdown Press the F2 key in the middle of a countdown to stop the process Restart the countdown by pressing the Start key F2 Alternatively press the Menu key F3 to go into the timer menu and...

Страница 27: ...on indicates that the countdown has expired To reset the timer press the Reset key F2 The display will look like this To repeat functions press the F2 key Note The C F function is disabled when the co...

Страница 28: ...imer temporarily overrides the auto off function until the countdown is completed or manually stopped If the meter is manually or automatically shut off the countdown alarm is automatically set to Dis...

Страница 29: ...ecall key or log key to choose the logging methods as auto or manual If auto logging is selected use the recall key or log key to set the time interval Its range is from 2 minutes to 60 minutes DATA T...

Страница 30: ...connection is established with a PC press the Select key F1 to download data from Meter to PC using HyperTerminal NOTE METER MUST BE TURNED ON PRIOR TO CONNECTING USB CABLE OR COMPUTER MAY NOT RECOGNI...

Страница 31: ...19 CALIBRATION REPORT SCREEN The calibration report will show the time and date along with results of the last user calibration 27...

Страница 32: ...0 CLEAR RESET SCREEN Press the F1 key to select the data you want to clear or reset For calibration logged data and reset all you will have to enter the password to proceed Default Password is 9900 28...

Страница 33: ...LECTROSTATIC DISCHARGE CLEAN WITH DAMP CLOTH Do not clean with abrasives or solvents Use mild detergents never immerse nor use excessive fluid 23 BATTERIES If there is no display when the thermometer...

Страница 34: ...indication Display Shows OPEN No RTD connected in the connector Display Shows Err If display shows this message other than during the field calibration mode please return the instrument for servicing...

Страница 35: ...rmometer 35426 60 Rubber Armour with Stand 35427 80 Handsfree Kit Two Magnets and a Strap 35427 85 General Purpose Probe immersion into liquids 08117 70 Penetration Probe meat semi soft materials 0811...

Страница 36: ...ide the warranty period or those resulting from product misuse or abuse may be invoiced to you 27 PRODUCT RETURN To limit charges and delays contact the seller or Manufacturer for authorization and sh...

Страница 37: ...tomers meet ISO FDA EPA GLPs cGMPs and other quality standards Conformity ISO IEC 17025 2005 accredited NIST Handbook 150 2000 Edition ANSI NCSL Z540 2 1997 NIST Technical Note 1297 ISO 9000 2000 Fast...

Страница 38: ...ur instrument We provide you with the documentation you need to meet your most stringent quality requirements for the control of inspection measuring and test equipment Certification includes certific...

Страница 39: ...send you an automatic reminder when it is time to recalibrate your instrument Are your calibration certificates good enough InnoCal surpasses the competition by providing the most complete certificat...

Страница 40: ...traceable standards Meter only Probe only System meter probe Four test points across range of instrument 0 100 165 230 C 4 32 446 770F 17000 04 17001 04 17002 04 InnoCal The Choice of Quality 1 866 In...

Страница 41: ...37...

Страница 42: ...38...

Страница 43: ...39...

Страница 44: ...r or authorized seller For more information on OAKTON Instruments Products please contact your nearest distributor or visit our web site listed below Oakton Instruments 625 E Bunker Court Vernon Hills...