Oakton T100WL Скачать руководство пользователя страница 13

 

- 13 - 

 

Using the same sample vial to calibrate and measure can eliminate the error caused by 

different vials, thus reaching higher accuracy. For example, add distilled water to 1

#

 vial for 

calibration and then add sample solution to 1

#

 vial for measurement. Note that the solution vial 

should be rinsed thoroughly when changing solutions. 

d) 

Zero Turbidity Error Reminder:

 0 NTU is the reference point for low-turbidity measurement. Due 

to the complexity of turbidity measurement, various inaccuracies resulting from instruments, 

solutions, vials, and operations can cause large errors at the 0 NTU point, which has a great impact 

on the measurement accuracy of low turbidity solutions. When the reading is displayed as 0.00 NTU, 

if a positive error occurs, a 0.0 NTU calibration solution can be used for verification test. If a 

negative error occurs, it cannot be verified. For this instrument, there is a Zero Turbidity Error 

Reminder function. If the negative error at the zero point exceeds the preset value, it will prompt 

users to recalibrate 0.0 NTU. At this time, users can press                to measure 1 or 2 more times. If 

the prompt appears continuously, please recalibrate 0.0 NTU. In addition, compared with other 

calibration solutions, The 0.0 NTU solution is more prone to deterioration and error. When the 0.0 

NTU solution is found to be turbid or flocculent, or if it is judged that it may be out of tolerance based 

on the experience of sample tests, please replace the 0.0 NTU solution in time according to the 

requirements in Section 4.1b. 

e) 

High turbidity calibration requirement: 

For turbidity measurement greater than 2 NTU, it is 

recommended to calibrate once a week, or to test a calibration solution close to the sample solution. 

If the error is large, the instrument needs to be recalibrated. 

f) 

The instrument does not automatically recognize the calibration solution. If the wrong solution is 

selected for calibration, the measurement will be completely wrong. In this case, it can be restored 

by recalibrating with the correct calibration solutions. 

g)  Place the instrument on a flat and level surface. Do not hold the instrument in your hands while 

operating. 

h)  If using Formazin standards for calibration, please note that the diluted Formazin standard is 

unstable. Please make sure to use the freshly made Formazin standard to ensure calibration 

accuracy. 

T URBIDIT Y  MEAS UREME NT  

5.1  Sample Vial Handling 

a)  6 sample vials are included in the test kit. The cap is marked with 1

#

 to 6

#

, and the bottom of the vial 

also has the same number. The number of the vial and the cap should always be the same. *Pay 

attention that 1

#

 and 2

#

 vials are only for low turbidity solution measurement. (< 2 NTU) 

b)  The vial has been rigorously cleaned and sterilized. They can be used directly for the first time. For 

subsequent uses, follow the steps below to perform a thorough cleaning. 

 

Clean the inside and outside of the sample vial holder 

with detergent → rinse with distilled 

water or deionized water multiple times → Rinse the vial twice with the sample solution → Pour 

the sample solution into the vial → Close the cap. 

Содержание T100WL

Страница 1: ...1 T100WL Portable Turbidity Meter 35635 10 Instruction Manual...

Страница 2: ...eparation for Calibration 10 4 2 Calibration Procedure Take 0 NTU and 20 NTU as an example 11 4 3 Notes for Calibration 12 5 Turbidity Measurement 13 5 1 Sample Vial Handling 13 5 2 Measurement Prepar...

Страница 3: ...parameter setting TruRead measurement mode automatically takes multiple consecutive readings and calculate their average minimum and maximum value as well as displaying each set of data in a brief te...

Страница 4: ...rmazin Solution 0 20 100 400 and 800 NTU FNU Detector Silicon photovoltaic Measuring Mode Normal measurement and TruRead measurement Data Storage 200 groups Data Output USB to PC Calibration Record Ca...

Страница 5: ...er when measuring Dust proof plug Take off the plug when measuring Housing Sample vial holder Display Calibration vials or sample vials Keypad Positioning mark Align the mark with the mark on the cali...

Страница 6: ...ofiber cloth Power adaptor 5V 1A Software flash drive Silicone oil 10 ml Sample vials 6 USB Cable under the meter User Manual Diagram 2 Measurement Mode Measurements and unit Operation Guide Battery i...

Страница 7: ...iagram 3 Calibration Mode Standard values Operation guide Finished calibration indicator Parameter Setup Menu Operation guide Data Recall Saved data numberings Saved data Operation guide Calibration M...

Страница 8: ...nt while it is being charged If charging the instrument when it is turned off Charging will be displayed and Charging is completed will be displayed after a full charge 3 6 Data Log Recall and Deletio...

Страница 9: ...surement mode Measurement Mode Storage Capacity Normal 200 sets TruRead 5 times measurement 100 sets TruRead 10 times measurement 61 sets TruRead 15 times measurement 44 sets TruRead 20 times measurem...

Страница 10: ...0 NTU calibration solution and close the vial cap Replace laboratory distilled water open the vial cap pour out the original solution add 1 2 distilled water cover the cap and shake the vial to rinse...

Страница 11: ...ake the solution even But sediment can still occur during tests and make the measurement unstable Users need to have rich experience with Formazin calibration solutions 4 2 Calibration Procedure Take...

Страница 12: ...n The calibration point can be verified after the calibration is completed If the calibration point has a large error enter the calibration mode and repeat the calibration For calibration point accura...

Страница 13: ...e replace the 0 0 NTU solution in time according to the requirements in Section 4 1b e High turbidity calibration requirement For turbidity measurement greater than 2 NTU it is recommended to calibrat...

Страница 14: ...number of continuous measurements is set to 20 times in parameter setting P2 a measured value will be displayed every 20 seconds Diagram 9 2 is the third measurement Diagram 9 3 is the display interf...

Страница 15: ...c for details c Mixing and Degassing Samples should not be vigorously shaken or vibrated It is recommended that users gently shake the sample vial to make solution evenly distributed Air bubbles in s...

Страница 16: ...lect in how soon the instrument will turn off automatically if there is no operation 10 minutes 20 minutes 30 minutes or off when selecting off the instrument can only be turned off by pressing the po...

Страница 17: ...er setting information Function buttons 7 2 Operation Key in Software Interface Refresh Port reset key When the instrument and computer are not connected click to connect SyncTime Sync time key click...

Страница 18: ...ed in the USB drive 7 4 Run Software Click Download key all data in the instrument memory will be sent to the computer When the program is running all measurement information will be sent to the compu...

Страница 19: ...are as in Diagram 13 Diagram 14 shows you how to replace the lamp Unscrew the lamp cover with a coin Use a tweezer to unplug the socket Use a tweezer to unscrew the lamp cover Take out the old tungste...

Страница 20: ...limited warranty Within the warranty period repair parts or product replacement will be made without charge due to malfunctions caused by manufacturing defects 9 2 The above limited warranty shall not...

Отзывы: