Symbol

Menu setting

Selection

Factory default

P1

P2

P3

P4

P5

P6

P7

Select pH buffer

solution

Set low

measurement alarm

Set high

measurement alarm

Select Automatic

Lock (HOLD)

Select backlight

Select temperature

unit

Restore to factory

default

USA – NIST

0 to 14.00 pH 

Off – On

Off – 1 – On

°C – °F

No – Yes

USA

0 to 14.00 pH

0

14.00

Off

1

°C

No

Setting the Parameters

When tester is turned off, long press    /MEAS key to enter 

parameter setting. Short press MODE/

△ 

key to switch from P1 to 

Notes

A. Select standard pH buffer solution (P1) There are two options of 

standard buffer solutions: USA series and NIST series.
B. High and Low Measurement Heads-Up Examples (P1 and P2) 

Alert when measured value ≤3.20 pH: 

Preset lowest value (P2) = 3.20 

pH, highest value (P3) = 14.00 pH, when measured value is ≤3.20 

pH (stability icon (

) displays on LCD), the red backlight appears on 

display. 

Alert when measured value ≥8.60 pH:

 Preset highest value (P3) = 8.60 

pH, lowest value (P2) = 0.00 pH, when measured value is ≥8.60 pH 

(stability icon (

) displays on LCD), the red backlight appears on 

display.

Alert when measured value ≤3.20 pH or ≥8.60 pH:

 Preset lowest 

value (P2) = 3.20 pH, highest value (P3) = 8.60 pH, when measured 

value is lower than 3.20 pH or higher than 8.60 pH (stability icon (

displays on LCD), the red backlight appears on display.
C. Automatic Lock (P4) 
Select “On” to activate auto lock function. When reading is stable for 

more than 10 seconds, the tester will lock the value automatically, 

and “

HOLD

” icon will appear on the bottom left of the screen. Press 

CAL/

   key to cancel HOLD on reading.

D. Backlight (P5) 
Select “Off” to turn off backlight function, “On” to turn on backlight 

unction, or “1” to have backlight last for 1 minute.
E. Temperature Unit (P6)
Select °C or °F; the factory default is °C.
F. Factory Default Setting (P7)
Select “Yes” to recover instrument calibration to the theoretical value 

(pH value in zero potential is 7.00 pH, slope is 100%), parameter 

setting returns to initial value. This function can be used when 

instrument does not work properly in calibration or measurement. 

Calibrate and measure again after resetting the unit to factory 

default status.

ORP Measurement

ORP stands for Oxidation-Reduction Potential. ORP is a measure of 
the cleanliness of the water and its ability to break down 
contaminants. Note that ORP sensor module 35634-58 or 35634-68 
must be purchased separately (see Ordering Information).

Once ORP sensor probe is attached to tester, press MODE/

△ 

key to 

enter ORP mode. Rinse the probe in distilled water and dry it. Dip 
the probe in sample solution, stir gently, and allow it to stand still 
until a stable reading is reached. Get readings after measurement 
stability icon (

) appears and remains on the screen.

pH 

ORP (mV)*

Temperature

Range
Resolution
Accuracy

–2.00 to 16.00 pH
0.01 pH

±0.01 pH ±1 digit

Calibration points

1, 2, or 3 points;

auto buffer recognition

Automatic 

temperature

compensation 

(ATC)

32 to 122°F(0 to 50°C)

Range

±1000 mV

Accuracy

±0.2% full-scale

Resolution

1 mV

Range
Resolution
Accuracy

32 to 122°F (0 to 50°C)
0.1°F/°C

±0.9°F (0.5°C)

Specifications

* Note that the ORP probe 35634-58 or 35634-68 must be purchased 

  separately (see Ordering Information).

Display: 

LCD with three-color backlight. Blue = measurement; 

Green = calibration; Red = alarm

Reading lock: 

HOLD icon

Power: 

four AAA batteries (included); >400 hours of continuous 

operation

Low-voltage warning:        

battery status icon flashes

Auto power-off: 

tester automatically turns off after 8 minutes of 

nonuse

IP rating: 

IP67 (waterproof), floats on water when sensor cap is on 

Dimensions (L x W x H): 

7" x 1.5" x 1. 5" (17.8 x 4 x 4 cm) 

Weight:

 4.7 oz (133 g)

Probe Replacement

Screw off the probe/sensor ring, unplug the probe, plug in the new 

replacement probe (pay attention to the probe’s position), and 

rescrew on the sensor ring.

Battery Installation

The tester uses four AAA batteries. Please install batteries accor ing 

to the following steps. Note the correct direction of battery installa-

tion: the positive side (+) of every single battery must face up. 

Incorrect installation of batteries will cause damage to the tester and 

create a potential hazard.

Warranty

We warrant this instrument to be free from defects in material and 

workmanship and agrees to repair or replace free of charge, at 

option of Oakton Instruments, any malfunctioned or damaged 

product attributable to responsibility of Oakton Instruments, for a 

period of two years from the delivery (a six-month limited warranty 

applies to sensors). This warranty does not apply to defects 

resulting from actions such as misuse (violation of the instructions in 

this manual or operations in the manner not specified in this 

manual), improper maintenance, and unauthorized repairs. 

Warranty period is the time limit to provide free service for the 

products purchased by customers, not the service life of the tester 

or probe.
Oakton Instruments reserves the right to update the information in 

this manual without giving notice in advance.

Model

Catalog

number

Product description

pH30

pHTestr 30 pocket pH tester

Replacement sensor for pHTestr 30

Optional probe/sensor for ORP testing

(single junction)

Optional probe/sensor for ORP testing

(double junction)

35634-30
35634-03
35634-58

35634-68

Ordering Information

Cap Lock

Clip

Battery Lid

Lift the clip to unlock

4×AAA

Batteries

Push the clip to open the battery lid

1. Open battery lid. 
2. Slide the battery cap along the direction of arrow. 
3. Open the battery cap. 
4. Insert the batteries

 (ALL POSITIVE SIDES FACING UP). 

5. Close the battery cap. 
6. Slide and lock the battery cap along the direction of arrow. 
7. Fit the tester’s cap while making sure to push all the way down. 

The tester’s waterproof design may be compromised if the cap is 
not fitted correctly. 

P2…P7. Short press CAL/

   key and parameter will flash, then short 

press MODE/

△ 

key to choose desired parameter. Short press 

CAL/

   key to confirm parameter selection. Long press    /MEAS key 

to return to measurement mode.