3 Bars indicates Battery is full (100%)
2 Bars indicates 50% of the battery life is left
1 Bar indicates 25% of the battery life is left
Blinking battery casing indicates the need to replace
batteries with fresh ones as specified by manufacturer
Electrode is not in contact with solution or electrode is
failing.
Replacement sensor is not connected properly to the
tester during sensor replacement
Measured pH value or temperature value (pHTestr30)
exceeds its specified maximum or minimum value
Blinking ‘ATC’, ‘Or’ or ‘Ur’ indicates that there is a short
or open circuit at the built in temperature sensor
Temperature calibration error of attempting to calibrate
tester to a value which is out of  range or under range
pH calibration error of attempting to confirm a calibration
value which is not within the specified calibration window

Self-Diagnostic Messages

Er

Er

Er

Er

Er.1

.1

.1

.1

.1

Low battery

indicator

Or / Ur

Or / Ur

Or / Ur

Or / Ur

Or / Ur

(Still)

Over range /
Under range

signal

AAAAATTTTTC / Or /

C / Or /

C / Or /

C / Or /

C / Or /

Ur 

Ur 

Ur 

Ur 

Ur (Blinking)

Er

Er

Er

Er

Er.0

.0

.0

.0

.0

Error

Message

Large Screen

pHTestr10

pHTestr20 pHTestr30

pHTestr

pHSpear

Testrs

10BNC

pH Range

-1.0 to 15.0 pH

-1.00 to 15.00 pH

Resolution

0.1 pH

0.01 pH

Relative Accuracy

0.1 pH

0.01 pH

Calibration Points

Up to 3 points

Up to 3 points

Buffer Set

USA - 4.0/7.0/10.0

USA - 4.01/7.00/10.01

Standard Selection

NIST - 4.0/6.9/9.2

NIST - 4.01/6.86/9.18

Calibration Window

+/-1.0 pH (pH 4.0

+/-1.00 pH (pH 4.01 & pH 10.01),

(USA Buffer Set

& pH 10.0),

+/-1.50 pH (pH 7.00)

Standard)

+/-1.5 pH (pH 7.0)

Calibration Window

+/-1.0 pH (pH 4.0

+/-1.00 pH (pH 4.01 & pH 9.18),

(NIST Buffer

& pH 9.2),

+/-1.25 pH (pH 6.86)

Set Standard)

+/-1.2 pH (pH 6.9)

Temperature

No

0-50.0°C or

Display

32.0-122.0°F

Temperature Compensation

ATC

MTC

Temperature  Resolution

No

0.1 °C / °F

Temperature  Accuracy

No

0.5 °C / 0.9 °F

No

Temperature Calibration

No

+/- (5°C /9 °F)

0-50.0°C or

Window

from default value

32.0-122.0°F (MTC)

Auto Off

After 8.5 minutes from last key press

User reset

Yes

Non Volatile Memory Backup

Yes

LCD Display

Dual

Power Requirement

4 x 1.5V “A 76” micro Alkaline Batteries

Battery life

More than 500 hrs

Operating Temperature

0 – 50 °C

Tester Dimensions

6.5 “L x 1.5”dia. (165 x 38 mm)

9.75 “L x

1.5”dia.

(247 x

38 mm)

Weight

3.25 oz (90 gm)

Penetrating electrode total length

Not applicable

91 mm

Shaft length

Not applicable

31 mm

Penetrating electrode

Not applicable

12 mm

upper diameter
Shaft diameter

Not applicable

7 mm

Tester Specifications

Temperature Calibration (Only for pHTestr 30)

From the measurement mode,

1. Press the HOLD/ENT button to bring the Testr to the ‘HOLD’ mode.

2. Press the CAL button for 3 seconds to switch to the °C or °F mode setting

selection screen. Pressing the CAL button continuously for 3 seconds allows you

to toggle in between the °C and °F mode setting selection screen.

3. Release the CAL button to confirm your mode selection and the display will go to

the temperature calibration mode with the upper display flashing. The upper

display shows the current measured temperature reading based on the last set

offset and the lower display shows the current measured temperature reading

based on factory default calibration.

4. Dip the Testr into a solution of known temperature and allow time for the in built

temperature sensor to stabilize.

5. Press the HOLD/ENT button to set the upper display to the temperature value of

the solution.

6. Once the new temperature setting is reached, the new value is automatically

confirmed and returns to the measurement mode if no button is pressed after 5

seconds.

Notes: To exit this program without confirming the calibration, press the CAL button before

the automatic confirmation takes place.

Continuous pressing of CAL button (about

3 seconds) to toggle between

temperature setting of degrees Celsius

and Farenheit

Release
CAL button
to start
entering
the new
temperature

During calibration, upper display will flash.

Press HOLD/ENT to increase offset up to
+5 °C (+9 °F) from default value.  It will

roll over to -5 °C (-9 °F) from default value

and start increasing again.

When the

temperature offset

value is reached,

the new value is

automatically

confirmed if no

key is pressed
for 5 seconds.

Figure 6: Temperature Calibration Sequence ( Only for pHTestr 30DJ)

Electrode Maintenance

1. Before measuring soil pH with the pHSpear, wet your soil sample with distilled

water and ensure that the soil is free of particulates. Even though the pHSpear is

ruggedized, hard surfaces such as stones and pebbles can still cause breakage.

2. Rinse the electrode with tap water or electrode storage solution after each

measurement. Care has to be taken not to damage the sensor’s glass electrode

especially while rinsing the pHSpear penetrating tip electrode.

3. In aggressive chemicals, dirty or viscous solutions, and solutions with heavy

metals or proteins, take readings quickly and rinse electrode immediately afterward.

For the pHSpear, the remnants of the semi solid samples on the penetrating

electrode can be removed by rubbing it with some table salt and then rinsing. Mild

detergent can be used to wash the penetrating electrode clean.

4. If possible, keep a small piece of paper or sponge in the electrode cap – moistened

with clean water or electrode storage solution (NOT de-ionized water) – and close

the cap over the electrode. For pHSpear, ensure that the electrode is kept soaked

in electrode storage solution or tap water via its protective cap.

Electrode Replacement

You can replace the electrode module at the fraction of  the cost of  a new Testr.

When the tester fails to calibrate or gives fluctuating readings in calibration standards, you

need to change the electrode.

1. With dry hands, grip the ribbed Testr collar with electrode facing you. Twist the

collar counter clockwise (see picture A). Save the ribbed tester collar and

O-ring for later use.

2. Pull the old electrode module away from the Testr.

3. Align the four tabs on the new module so that they match the four slots on the

Testr (see picture B).

4. Gently push the module onto the slots to sit it in position. Push the smaller

O-ring fully onto the new electrode module. Push the collar over the module

and thread it into place by firmly twisting clockwise.

Note: It is necessary that you recalibrate your Testr prior to measurement after an electrode

replacement.

Applications

Water quality testing • pools • spas • aquariums • aquaculture • hydroponics • ecology

studies • water and wastewater treatment • boilers • steam generators • car washes

• sanitation plants • labs • food sectors and more!

Rlbbed collar

Rotate collar

away from you

Insert Electrode Module

Electrode Module

Large

O-ring

Small O-ring

Small Tab

Large Tab

68X068042  Jun 10  Rev 7

Figure 8: Example of electrode module

fitting alignment

Figure 7: Removal of collar

from tester

Changing Batteries

1. Open battery compartment lid (with attached lanyard loop).

2. Remove old batteries; replace with fresh ones. Note polarity

Accessories

Item

Item

Item

Item

Item

Eutec

Eutec

Eutec

Eutec

Eutech Instr

h Instr

h Instr

h Instr

h Instruments

uments

uments

uments

uments

Oakton Instr

Oakton Instr

Oakton Instr

Oakton Instr

Oakton Instruments

uments

uments

uments

uments

Or

Or

Or

Or

Order Code

der Code

der Code

der Code

der Code

Or

Or

Or

Or

Order Code

der Code

der Code

der Code

der Code

pHTestr 10,20,30 replacement sensor

PHSENSOR03DJ

35624-38

pHSpear replacement sensor

PHSENSOR04

35634-50

Carrying Pouch pH Testr 10, 20, 30

ECPOUCH01

35624-45

Warranty

The waterproof Testrs are warranted to be free from manufacturing defects for 1 year and

electrode module for 6 months, unless otherwise specified. If repair, adjustment or replacement

is necessary and has not been the result of abuse or misuse within the time period specified,

please contact your local distributor for assistance.

Return of Items

Authorization must be obtained from your distributor before returning items for any reason.

When applying for authorization, please include information regarding the reason the item(s)

are to be returned.

Note: We reserve the right to make improvements in design, construction and appearance of

products without notice. Prices are subject to change without notice.

 

 

 

Содержание pHTestr 30

Страница 1: ...ion manual pH Buffer Set Selection Your Testr features USA pH 4 01 pH 7 00 and pH 10 01 or NIST pH 4 01 pH 6 86 and pH 9 18 standards Select either one to suit your requirements 1 While pressing the HOLD ENT button switch on the Testr by pressing the ON OFF button 2 Release the HOLD ENT button The display will flash either USA or NIST 3 Press CAL button to toggle between the two buffer set standar...

Страница 2: ...l for sample contact CAUTION Testing dry samples is not accurate and can lead to sensor damage or breakage Soils must be wet and free of particulates that may scratch the glass sensor Excessive force into dry samples can cause glass breakage 3 Note the pH value or press HOLD ENT button to freeze the reading To release the reading press HOLD ENT again 4 Press ON OFF to turn off Testr If you do not ...

Страница 3: ...lower display shows the last set temperature offset 4 Press the HOLD ENT button to set the upper display to the temperature value of your sample 5 Once the setting is reached release the HOLD ENT button The new value is automatically confirmed and returns to the measurement mode if no button is pressed after 5 seconds Notes To exit this program without confirming the calibration press the CAL butt...

Страница 4: ...ng the new temperature During calibration upper display will flash Press HOLD ENT to increase offset up to 5 C 9 F from default value It will roll over to 5 C 9 F from default value and start increasing again When the temperature offset value is reached the new value is automatically confirmed if no key is pressed for 5 seconds Figure 6 Temperature Calibration Sequence Only for pHTestr 30DJ Electr...

Страница 5: ...ost of a new Testr When the tester fails to calibrate or gives fluctuating readings in calibration standards you need to change the electrode 1 With dry hands grip the ribbed Testr collar with electrode facing you Twist the collar counter clockwise see picture A Save the ribbed tester collar and O ring for later use 2 Pull the old electrode module away from the Testr 3 Align the four tabs on the n...

Страница 6: ...ft length Not applicable 31 mm Penetrating electrode Not applicable 12 mm upper diameter Shaft diameter Not applicable 7 mm Tester Specifications 68X068042 Jun 10 Rev 7 Accessories Item Item Item Item Item Eutec Eutec Eutec Eutec Eutech Instr h Instr h Instr h Instr h Instruments uments uments uments uments Oakton Instr Oakton Instr Oakton Instr Oakton Instr Oakton Instruments uments uments uments...

Отзывы: