background image

 

- 5 - 

Clear stored data 

The meter can store 50 groups of data. Once the storage is full, if user 

 

presses            again, the 

FUL

 icon will be displayed on LCD (see Figure-6). 

 

The storage needs to be cleared in parameter setting P4 so as to keep storing 

 

data. The numberings in 

M+ 

indicates the total number of stored data; The 

 

numberings in RM indicates the number of stored data in current mode (ph/mV). 

 

For example, if 20 groups of data are stored in pH mode, and 10 groups in mV 

 

mode, then the numberings for M+ will be “

M+30

” in both pH and mV mode; 

 

the numberings for RM in pH mode would be “

RM20

”, in mV mode “

RM10

”.

 

 

 

4.6 

Max/Min Function

 

Use  this  function  to  display  the  maximum  or  minimum  records  out  of  groups  of  measuring  data.  In 
measurement  mode,  press          ,  LCD  will  display  flashing  icons  of  “

MAX/MIN

”,  indicating  the  meter  has 

entered  the  mode  of  recording  maximum  and  minimum  data.  After  testing,  press          again,  LCD  will 
alternately display the maximum and minimum value in the past measurements. Press            to go back to 
regular measurement mode.   

 

 

pH Measurement

 

5.1 

Preparation

 

Install the electrode holder 

The electrode holder is composed of a metal stand base, a supporting pole, and an electrode clip. First, install 
the metal stand base to the meter (screw it on). Please note that the metal base can be installed on either side 
of the meter. Then install the supporting bar on the metal stand base, and then install the electrode clip on the 
supporting bar.

 

 

Connect the power adapter 

Plug the power adapter tightly into the DC9V socket. The power’s voltage should meet the requirements listed 
on the power adapter.

 

 

Things needed in addition to 

what’s included in the box

 

A clean cup, distilled water (8-16oz), and tissue papers for rinsing and drying the probe.

 

 

5.2 

pH Electrode Information

 

pH Electrode 

The meter comes with the 35419-31, 3-in-1 combination electrode with a built-in temperature sensor, which 
enables the automatic temperature compensation. This electrode is suitable for regular water solutions’ pH 
testing. Please refer to Section 7 for ideal pH electrodes to use for other specific applications. The electrode’s 
housing adopts polycarbonate materials, which is resistant to shocks and corrosion. A 3M KCL storage bottle 
comes with the electrode, which is for storing the electrode when not in use to keep the sensitivity of the probe.

 

Figure - 6

 

Содержание PH550

Страница 1: ...PH550 Benchtop pH Meter Instruction Manual Oakton Instruments www 4oakton com Manual Version 3 2...

Страница 2: ...4 5 Data storage recall and deletion 4 4 6 Max Min Function 5 5 pH Measurement 5 5 1 Preparation 5 5 2 pH Electrode Information 5 5 3 Information regarding Calibration 6 5 4 pH Meter Calibration 8 5 5...

Страница 3: ...g technology intelligently improves the meter s response time and measuring accuracy with stable reading display mode available The kit comes with a 3 in 1 combination pH Electrode measuring pH and te...

Страница 4: ...bration icons and numberings 3 Measurement unit 4 Temperature and reminder icons 5 The pH unit displayed during calibration 6 The pH value displayed during calibration numberings for storage and recal...

Страница 5: ...on and go back to measurement mode Short Press In calibration mode press to conduct calibration In parameter setting press to confirm selection Short Press Store measuring data Long Press Recall store...

Страница 6: ...deletion Storage When the measurement is stable press key the meter displays M icon and storage numberings on LCD and stores the measuring data In Figure 4 the meter stores the first group of the meas...

Страница 7: ...ectrode holder The electrode holder is composed of a metal stand base a supporting pole and an electrode clip First install the metal stand base to the meter screw it on Please note that the metal bas...

Страница 8: ...water Gently stir the electrode for a few seconds after it s dipped into the test solution and then let it stand still Wait for the stable measurement a smiley face appears and stays and then record...

Страница 9: ...n of electrodes and the requirement of the accuracy For High Accuracy meaustements 0 02pH the meter should be calibrated before test every time For ordinary accuracy measurements 0 1pH once calibrated...

Страница 10: ...d water dry it and dip it into pH 4 00 buffer solution Stir the solution gently and let stand still in the buffer solution until a stable reading is reached The meter s display will show the recogniti...

Страница 11: ...e purified water A small addition of KCL solution will help with the measurement time 2 Purified water such as distilled water and deionized water are recommended for only rinsing the probe for the be...

Страница 12: ...bration and measurement the meter has a self diagnosis function Please see table 5 Table 5 Self diagnosis Information Display Icons Contents How to fix Wrong pH buffer solution or the buffer solution...

Страница 13: ...menu from P1 to P2 P5 For details please see Table 7 Table 7 pH Parameter Setting P1 Select pH buffer series USA NIS 1 Long press to enter P1 2 Press USA flashes press to select USA or NIS press to co...

Страница 14: ...calibration accuracy the pH of the standard buffer solution must be reliable The buffer solution should be refreshed often especially after heavy use We recommend the Oakton buffer assortment pack 059...

Страница 15: ...easuring value use ORP standard solution to test mV value and see whether ORP electrode or meter works properly Table 9 is the data of standard Zobell ORP solution for 222 mV Table 9 C 10 15 20 25 30...

Страница 16: ...lfunctioned or damaged product attributable to responsibility of OAKTON INSTRUMENTS for a period of THREE YEARS SIX MONTHS for the probe from the delivery This limited warranty does not cover any dama...

Отзывы: