background image

 

 PCTSTestr

 50 Pocket Tester

Measurement Parameter Setting

1. Press ON/OFF ( ) to power on the tester.
2.  Press MENU/

v

 

to enter setup window. Press HOLD/

   

to select Measure. The display shows pH, Cond, TDS  

and Salinity.

3.  Scroll down by pressing MENU/

v

 

to toggle between pH, 

Cond, TDS and Salinity. Press HOLD/

  

to select pH.

4.  The display shows the selected parameter with a  .

pH Buffer Set Selection

PCTSTestr 50 Pocket Tester features USA (pH 4.01, pH 7.00 

and pH 10.01) or NIST (pH 4.01, pH 6.86, and pH 9.18)  

standards. Select either one to suit your requirements.
1.  Press MENU/

v

 

to enter setup window. Press HOLD/

   

to select Settings. The display shows Buffer, TDS Factor  

and Backlight.

Settings
Temp Set
Reset

Cond
TDS 
Salinity

hold

pH

menu

hold

pH

Measure

2.   Press  HOLD/

   

to select Buffer. Display shows USA and NIST.

3.  Press HOLD/

   

to select USA or scroll down by pressing       

MENU/

to toggle between the two buffer standards.

4.  The display shows the selected buffer standard with a  .

TDS Factor Setting

1.  Press MENU/

v

 

to enter setup window. Scroll down by  

pressing MENU/

to select Settings.

2.  Press HOLD/

   

to select Settings. The display shows  

Buffer, TDS Factor and Backlight.

3.  Scroll down by pressing MENU/

to toggle between the 

Buffer, TDS Factor and Backlight. Press HOLD/

   

to select 

the TDS Factor.

4.  Press HOLD/

   

to select the default TDS factory setting or  

MENU/

to adjust the setting.

5.   Press  HOLD/

   

to confirm the selection of the setting. The 

display shows the selected value (TDS factor) with a  .

USA

Measure

Temp Set
Reset

TDS Factor
Backlight

hold

NIST

menu

hold

Buffer

Settings

hold

USA

Adjust  

TDS Factor

0.71

Measure

Temp Set
Reset

Buffer

Backlight

hold

menu

hold

TDS Factor

Settings

hold

0.66

Buffer
TDS Factor

Backlight

Backlight
ON
OFF

Measure

Temp Set
Reset

hold

menu

hold

Settings

hold

OFF

°F

Measure
Settings

Reset

Temp Cal
Temp Coeff

hold

menu

hold

Set °C/°F

Temp Set

hold

°C

°C

Temp Cal
23 4 °C
23 4 °C

Measure
Settings

Reset

Set °C/°F

Temp Coeff

hold

menu

hold

Temp Cal

Temp Set

hold

2.00

Getting Started

The PCTSTestr 50 Pocket Tester has been factory calibrated 

and usually works well out of the box. However, after 

extended periods of non-use, it is best to remove the sensor 

cap and soak the sensor in warm tap water for 10 minutes or 

so. Prior to taking the measurements, periodic calibration 

with certified standards is recommended for best accuracy.

Backlight Settings

1.  Press MENU/

v

 

to enter setup window. Scroll down by  

pressing MENU/

v

 

to select Settings. 

2.  Press HOLD/

   

to select Settings. The display shows  

Buffer, TDS Factor and Backlight.

3.  Scroll down by pressing MENU/

v

 

to toggle between  

Buffer, TDS Factor and Backlight. Press HOLD/

   

to select 

Backlight.

4.  The display shows ON and OFF. Scroll down by pressing        

MENU/

v

 

to toggle between ON and OFF. Backlight ON 

increases readability in low-light conditions.

5.  Press HOLD/

   

to select the desired backlight option. The 

display shows the selected backlight option with a  .

Temperature Settings

1.  Press MENU/

v

 

to enter setup window. Scroll down by 

pressing MENU/

v

 

to select Temp Set. Press HOLD/

 

  

to select Temp Set. The display shows Set °C/°F, Temp Cal 

and Temp Coeff.

2.  Press HOLD/

   

to select Set °C/°F. Scroll down by pressing  

MENU/

v

 

to toggle between °C and °F. 

3.  Press HOLD/

   

to select temperature unit. The display 

shows the selected temperature setting with a  .

Temperature Calibration 

1.  Press MENU/

v

 

to enter setup window. Scroll down by 

pressing MENU/

v

 

to select Temp Set.

2.  Press HOLD/

   

to select Temp Set. The display shows  

Set °C/°F, Temp Cal and Temp Coeff. 

3.  Scroll down by pressing MENU/

v

 

to toggle between  

Set °C/°F, Temp Cal and Temp Coeff. Press HOLD/

   

to select Temp Cal.

4.  The lower display shows the current measured  

temperature reading based on the last set offset and the 

upper display shows the current measured temperature 

reading based on factory default calibration.

5.  Dip the tester into a solution of known temperature and 

allow time for the built-in temperature sensor to stabilize.

6.  Press MENU/

v

 

to adjust the temperature value or press 

the HOLD/

   

to confirm the calibrated value as the new 

temperature value of the solution.

Note: To exit this program without confirming the  

calibration, press CAL/ESC. 

Applications

• Agriculture

• Aquaculture

•  Aquariums and 

fish farms

•  Boiler blow-

down

• Car washes

•  Drinking water 

• Ecology 

•  Electroplating 

rinse tanks

• Food sectors

•  Hydroponics

• Labs

• Printing industry

• Swimming pools

•  Verification of 

reverse osmosis 
system operation

•  Water and waste

-

water treatment

Operating Instructions

 

1

2

3

Calibration for Conductivity, TDS, or Salinity

For best results, periodic calibration with an accurate  

standard is recommended prior to measurement. Use the 

calibration standard value that is close to your intended  

sample value. The tester will retain one calibration value in 

each mode (conductivity, TDS, salinity) when the instrument 

is powered off. The conductivity value can be calibrated 

automatically or manually, while the TDS & salinity values 

require manual calibration. The tester will begin in the  

measurement mode that was used when it was powered  

off. See “Measurement Parameter Setting” to change the 

desired parameter.

Automatic Calibration for Conductivity

1. Remove the cap and press ON/OFF ( ) to power on.
2. Dip the sensor in at least 30 mm of calibration standard.
3. Stir gently and press CAL/ESC to begin the calibration.
4.  The display will show CAL followed by the default value. 

CAL is indicated on the display during calibration mode.

5.  If the reading is within the calibration range of the automati-

cally recognized standards; 80 (84 μS/cm), 1410 (1413 μS/cm), 

or 12.90 (12.88 mS/cm), the   icon is displayed when the 

automatic calibration standard value has been detected.

6.   Press  HOLD/

   

to accept the auto conductivity standard 

and finish the calibration. 

7. Display returns to Measurement window. 

Manual Calibration

When the conductivity reading is outside calibration range 

 

of the automatic conductivity standards or when TDS or 

salinity is used, the tester will require manual adjustment.
1.  Repeat steps 1 to 4 from “Automatic Calibration for  

Conductivity”.

2.   Press  MENU/

v

 

to manually adjust the value to the  

desired reading.

Note: The adjustment will decrease only, however the  

adjustment will eventually cycle to the highest available  

value after decreasing by 40% of the initial value.
3.   Press  HOLD/

   

to accept and finish the calibration when  

the desired value is selected. 

Note:  To abort calibration, press CAL/ESC to escape.

µS/cm 

844 

23.5°C

µS/cm 

844 

1413

Place the tester in 
the calibration STD 
to start or press   
     to escape.

cal

esc

cal

esc

CAL

CAL

COND

Factory  

Reset

User Reset

Yes

No

Measure
Settings
Temp Set

hold

menu

hold

Reset

hold

4.  Once the calibration is finished and user has accepted  

the changes, measurement window will now show the  

calibrated reading.

Note: The auto conductivity standards are 84 μS/cm,  

1413 μS/cm & 12.88 mS/cm.

pH Calibration

Calibration should be done regularly, recommended once a 

week. Calibrate up to three points using either the USA or 

the NIST buffer set standards.
1. Press ON/OFF ( ) to power on the tester if needed.
2.  Dip electrode about 2 cm to 3 cm into the pH standard  

buffer solution.

3.  Stir gently and press CAL/ESC to enter calibration mode. The 

CAL indicator will be displayed. The upper display will show 

the measured reading based on the last calibration while the 

lower display will indicate the pH standard buffer solution.

Note: To abort calibration, press CAL/ESC to escape.
4.  Allow about 2 minutes for the tester reading to stabilize. 

The timer icon blinks during this time. Once the reading  

is stabilized, the timer stops blinking. Automatic  

confirmation happens when the buffer is found and  

the display returned to measurement window with  

reading calibrated to pH standard buffer solution.

5.  Repeat with other buffers if necessary. Rinse electrode 

before dipping into next buffer.

Note:

 The calibration mode allows you to perform up to 

three calibration points. Calibration is automatically con-

firmed with the buffer identification. No user interaction is 

required after starting the calibration by pressing CAL/ESC.    

Measurement

1. Press ON/OFF ( ) to power on the tester if needed.
2.  Dip the electrode in about 2 cm to 3 cm into the test  

solution. Stir and let the reading stabilize. The timer icon 

will blink during this time. Once the reading is stabilized, 

the timer stops blinking and   will appear to indicate the 

stability of the reading.

    CAUTION:  Testing dry samples is not accurate 

and can lead to sensor damage or 

breakage. Soils must be wet and 

free of particulates that may scratch 

the glass sensor. Excessive force 

into dry samples can cause glass 

breakage.

3.  Note the value or press HOLD/

   

to freeze the reading.  

To release the reading, press HOLD/

   

again.

4.  Press ON/OFF ( ) for 3 seconds to turn off tester. If you  

do not press a button for 8.5 minutes, the tester will  

automatically shut off to conserve batteries.

User Reset

Reset to the user’s default settings by using the User Reset 

function. Buffer selection and user temperature calibration 

are not affected by the user reset function.
1.  Press MENU/

v

 

to enter setup window. Scroll down by 

pressing MENU/

v

 

to select Reset. Press HOLD/

   

to select 

Reset. The display shows User Reset and Factory Reset.

2.  Press HOLD/

   

to select User Reset. 

3.  The display automatically shows No and Yes. Scroll down 

by pressing MENU/

v

 

to toggle between No and Yes. 

4.  Press HOLD/

   

to confirm either No or Yes. The display 

shows the User Reset option with a  . 

Temperature Coefficient 

1.   Pres MENU/

v

  

to enter setup window. Scroll down by 

pressing MENU/

v

 

to select Temp Set. 

2.  Press HOLD/

   

to select Temp Set. The display shows  

Set °C/°F, Temp Cal and Temp Coeff.

3.  Scroll down by pressing MENU/

v

 

to toggle between  

Set °C/°F, Temp Cal and Temp Coeff.

4.  Press HOLD/

   

to select Temp Coeff or MENU/

v

 

to  

adjust the Temp Coeff.

5.  Press HOLD/

   

to confirm the Temp Coeff value. The  

new value is automatically confirmed with a   .

pH 

7.01 

25.5°C

pH

7.45 
7.01

Place the tester 
in the buffer to 
start  or  press                 
to escape.

cal

esc

cal

esc

CAL

CAL

pH

µS/cm 

844 
820

µS/cm 

844 

1413

Place the tester in 
the calibration STD 
to start or press   
     to escape.

cal

esc

cal

esc

CAL

CAL

CAL

µS/cm 

1413 

23.5°C

COND

4

5

6

Set °C/°F
Temp Cal

hold

Adjust  

Temp Coeff 

2.00

menu

 2.00

hold

Temp Coeff

hold

Measure
Settings
 
Reset

Temp Set

 

pH

  
  4.01

 

   

25.5°C

pH

Key Info   

Scroll 

Accept 

EXIT 

   

MENU 

HOLD 

CAL 

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   