SETUP FUNCTIONS
Use the setup feature to customize your instrument operation. First,
select the appropriate measurement mode you wish to adjust by
pressing
until the desired units are displayed (i.e. pH, mV,
Conductivity, or TDS). During measurement, press and hold
for 5
seconds to enter SETUP mode of the parameter being measured.
Press
or
to change programs or change options.
Press
to select the program or confirm selection.
Press
to go back one level or return to measurement mode.
9.1 1.0 CAL (Calibration)
Press
to view stored calibration points.
9.2 2.0 ELE (Electrode Information)
Press
to view mV offset (OFS)—mV and pH modes.
Press
to view % slope (SLP)—pH mode.
Press
to view cell efficiency in each range—conductivity
and TDS modes.
TIP:
When multiple pH slopes exist, the slope that is available to view
will depend on the measured value before entering SETUP mode.
For example, if pH calibration values pH 4.01, 7.00 and 10.01 were
completed, there will be two slope values; one that corresponds to pH
values below 7.00 and another that corresponds to pH values above
7.00. If the measured pH value was 5.23 when setup mode was
entered, the pH slope that corresponds to values lower than 7.00 will
12
be available to view. If no calibration is stored, the offset will be 0.0
mV and the slope is 100%.