3.  CALIBRATION 

3.1  pH Calibration 

The 5+ and 6+ meter is capable of calibrating up to 5 pH values using the USA or NIST 
buffer groups, or 2 pH values using the Low Ionic buffer group. All new calibration 
values will automatically override existing calibration data. The non-volatile memory will 
retain calibration values upon power shut-off, and battery removal.  

For best results, periodic calibration with known accurate standards is recommended. 
Calibrate with standards that bracket your intended measuring range while including a 
neutral standard (i.e. pH 7.00, 6.86, 6.97). For example, if you expect to measure 
samples from pH 6.2 to pH 9.5, calibration using 4.01, 7.00, and 10.01 will work well.  

To eliminate temperature errors associated with the pH electrode, attach the automatic 
temperature compensation (ATC) probe for best accuracy. Without temperature 
compensation, pH accuracy will worsen as samples deviate from 25 ºC and pH 7.  

Always rinse electrodes with clean water before and after each calibration or 
sample measurement to avoid cross-contamination. 

The following calibration standards are automatically recognized; 
 

Buffer Group 

Available pH Calibration Values 

USA

 (



)

 

1.68, 4.01, 7.00*, 10.01, 12.45 

NIST

 (



)

 

1.68, 4.01, 6.86*, 9.18, 12.45 

Pb 

(



)

 

4.10, 6.97 

*These are the only values suitable for a one-point calibration 

See 

Section 3.1—Changing the pH Buffer Group

 

to select a different buffer group. 

 

Always  use  fresh  pH  buffer  solutions  for  calibration.  Do  not  reuse  buffer  solutions  as 

they  change  with  prolonged  exposure  to  air  (especially  pH  10  buffer)  resulting  in 

decreased  measurement  accuracy.  Promptly  seal  containers  and  store  solutions  in  a 

dark, dry, cool environment. 

 

 

6

Содержание Ion 6+

Страница 1: ...Instruction Manual pH 5 pH C pH 6 pH C mV Ion 6 pH C mV Ion Technology Made Easy 68X576801 Rev 0 07 2011...

Страница 2: ......

Страница 3: ...edure 7 Changing the pH Buffer Group 8 Resetting User Calibrated Values 9 3 2 Ion Calibration Ion 6 only 10 Ion Calibration Procedure Ion 6 only 10 3 3 Millivolt mV Calibration 6 only 12 3 4 Temperatu...

Страница 4: ...es meters advance our popular 5 6 series meters that were introduced in 1998 Each meter includes alkaline AAA batteries a rubber armor stand and instruction manual Please refer to Section 7 Replacemen...

Страница 5: ...s the meter on and off Upon power on the meter automatically begins in the measurement mode that was last used MODE Selects measurement modes pH mV Ion Ion6 only Temperature With meter off press and h...

Страница 6: ...reset menu while powering on Increase value or scroll up in Setup or Cal modes Decrease value or scroll down in Setup or Cal modes HOLD Freezes measured reading indicated by HO Press again to resume l...

Страница 7: ...om the top of meter before pushing the bottom edges of meter into position Tilt the stand at the back of meter for table top usage as desired Inserting New Batteries The LO annunciator alerts you when...

Страница 8: ...ess All LCD segments will display momentarily during the self diagnostic test before scrolling the model number software revision pH buffer group and number of pH calibration points selected before re...

Страница 9: ...re errors associated with the pH electrode attach the automatic temperature compensation ATC probe for best accuracy Without temperature compensation pH accuracy will worsen as samples deviate from 25...

Страница 10: ...display briefly The pH value is automatically adjusted to the buffer value shown from your selected buffer group The example below shows a successful pH 7 00 calibration at 25 C 5 For a one point cali...

Страница 11: ...s to confirm the pH buffer group Note If is selected measurement mode will begin If or is selected proceed to step 5 5 Press to toggle between the desired number of calibration points Pt Note During p...

Страница 12: ...alibration depending on the selected mode Factory Reset of all calibration values and user settings to factory default settings For calibration reset measure in the mode that you want to reset prior t...

Страница 13: ...r each calibration or sample measurement to avoid cross contamination The Ion 6 will show when ion calibration is required Ion Calibration Procedure Ion 6 only 1 Turn on the meter and select ion mode...

Страница 14: ...ibration has been performed the meter automatically completes the calibration 8 A successful calibration will show slope followed by the mV decade value the display will cease blinking and begin ion m...

Страница 15: ...tandard in which adjustments are made in an attempt to match the ORP value however results are often difficult to reproduce 1 Turn on the meter and select mV mode by pressing if necessary Dip the ORP...

Страница 16: ...accurate 1 Turn on the meter and select C mode by pressing if necessary Connect the temperature probe and place it into a solution with a known accurate temperature such as a constant temperature bat...

Страница 17: ...ation ATC is recommended for best accuracy 1 Turn on the meter and select C mode by pressing if necessary Disconnect the temperature probe 2 Press to begin temperature calibration mode calibrate will...

Страница 18: ...desired mode of operation pH mV Ion or Temperature 3 Dip both probes gently into an aqueous test sample swirl or stir gently and allow the reading to stabilize Holding Releasing a Reading To freeze o...

Страница 19: ...olerance Use new calibration solution recalibrate Ensure correct pH buffer group was selected Ion calibration exited after 1 point calibration Perform at least 2 point ion calibration before exiting I...

Страница 20: ...45 USA 1 68 4 01 6 86 9 18 12 45 NIST 4 10 6 97 Pb Temperature Range 0 0 to 100 0 o C Resolution 0 1 o C Accuracy 0 5 o C Compensation Automatic Manual 0 to 100 o C Millivolt Range 1000 to 1000 mV Mi...

Страница 21: ...ng case ECPH601PLUSK 01X245028 pH 6 with pH and ATC probes and solutions in hard carrying case ECPH602PLUSK 01X245026 pH 6 with pH ATC probe and solutions in hard carrying case ECPH603PLUSK 01X245027...

Страница 22: ...e 1 pint ECBU4BT 00654 00 pH 4 01 buffer sachets 20 mL x 20 pcs ECBU4BS 35653 01 pH 6 86 buffer solution 480 mL bottle ECBU686BT 00654 03 pH 7 00 buffer solution 480 mL bottle 1 pint ECBU7BT 00654 04...

Страница 23: ...inadequate maintenance by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products 9 RETURN OF ITEMS Authorization must be obtained from our Cu...

Страница 24: ...sites listed below Oakton Instruments 625 E Bunker Court Vernon Hills IL 60061 USA Tel 1 888 462 5866 Fax 1 847 247 2984 info 4oakton com www 4oakton com Eutech Instruments Pte Ltd Blk 55 Ayer Rajah C...

Отзывы: