Figure 1:

 

Removing battery cover  

to change batteries

Battery 

cover

Step 2

Lift and push up

Step 1

Lift and hold

Before First Use:

 

Precipitation of the sensor electrolyte may cause white crystals to 

form around the cap. This is normal and does not affect the tester’s performance.

Conditioning:

 

Condition the tester before first use. Remove the transparent cap  

and fill with tap water to a height of 20 mm. Recap the tester and soak* sensor for  
one hour. This will activate the sensor and dissolve any crystals present. Rinse the 
cap after use.

Calibration:

 

This tester allows up to 3-point calibration. Start with pH 7.0 buffer 

 

• Press 

on/off

 button to power on. Immerse sensor in about 20 mm of buffer and 

stir gently. Wait for displayed value to stabilize at or near pH 7.0 • Press 

cal

 button  

to enter calibration sequence. 

CAL

 will flash momentarily and then show a flashing 

default reading • To complete the first calibration point, press 

hold/ent

 button to 

confirm (

Ent

). If necessary, proceed to next buffer value (pH 4.0 or pH 10.0) and 

repeat the calibration procedure to obtain your second and third calibration point 

•  To  abort  calibration,  press 

cal

  button  to  escape  (

ESC

).  You  have  successfully 

calibrated the instrument. 

Note:

 If no button is pressed for 15 seconds, tester will automatically confirm calibration.

pH Testing:

 

Remove cap and press 

on/off

 button to turn on the tester • Dip the 

sensor in at least 20 mm of test solution • Stir once and let the reading stabilize. 

 

Note the pH value • Press 

hold/ent

 button to hold (

HO

) the reading. Annunciator 

will flash to indicate reading is in hold mode. Press 

hold/ent 

again to cancel (

HC

the hold mode • Press the 

on/off

 button to shut tester off.

Note:

 Tester automatically shuts off after 8.5 minutes of non-use to conserve batteries.

Maintenance:

 

Rinse  the  sensor  in  tap  water  after  use.  In  harsh  samples,  take 

readings quickly, then rinse sensor immediately afterwards with de-ionized water 

 

to remove residues and prevent electrode contamination.

Useful Notes:

 

 

To avoid cross-contamination, rinse sensor between samples and 

buffer with de-ionized water • To maintain tester accuracy, use tester regularly or 
calibrate at regular intervals as required • For long-term storage, fill the cap with 

 

20 mm of storage solution (NOT de-ionized water), and replace cap*. This will keep 

the sensor conditioned and ready for next use.

* The tester cap is not leak-proof. Please keep tester upright during soaking.

Operating Instructions

EcoTestr pH 2 Waterproof Pocket Tester

Changing Batteries:

 

To remove batteries, lift up front battery cover and hold in 

position before lifting two sides of the pocket clip (Figure 1). Remove old batteries 

and replace with fresh ones. Note polarity as shown in battery compartment. No 

re-calibration is required after battery change since tester’s non-volatile memory 

function retains calibration.

Error Messages:

 bAt

 – Batteries are weak and need replacement. 

Err

 – Wrong or 

bad buffer value (out of range) and/or the sensor is failing. Use fresh buffer. 

Or/Ur

 – 

Over or Under range signal or sensor is contaminated/damaged/not fully immersed. 

Reset Option:

 Reset option allows you to restore the calibration back to factory 

default settings • Press 

on/off

 to turn off tester • Press and hold 

hold/ent

 button and  

then switch on using the 

on/off

 button. The display shows a flashing 

rSt

 (reset)  

•  Press 

hold/ent 

button  again  to  confirm  (

Ent

)  reset  or  press  the 

cal

  button  to 

escape (

ESC

) reset option.

Warranty:

 This tester is supplied with a warranty against manufacturing defects for 

a period of ONE year from date of purchase.

Accessories:

Eutech  

Order Code

Oakton  

Order Code

Description

ECBU4BT

00654-00

pH 4.01 buffer solution, 480 ml bottle

ECBU7BT

00654-04

pH 7.00 buffer solution, 480 ml bottle

ECBU10BT

00654-08

pH 10.01 buffer solution, 480 ml bottle

ECRE005

00653-04

Storage solution for pH sensor, 480 ml bottle

ECDPCBT

00653-06

Protein remover solution, 480 ml bottle