Use the System Setup to customize operation of your 2700 series meter. Press
SETUP
from the
measurement screen and then press
ENTER
when
SYSTEM
is highlighted to access these settings.
The following settings can be customized for each model:
n
Stability
To aid with measurement and calibration, your meter has a unique feature that provides an
obvious visual indication when the reading becomes stable. If the Stability feature is
enabled
,
the digits on the primary reading will be faded when unstable, becoming solid when stable.
If this feature is
disabled
, the primary reading will always be solid.
n
Stability Criteria
The 2700 allows adjustment of the Stable indicator which relates to how fast and frequent it
appears. There are three levels of adjustment;
SLOW
,
MEDIUM
, &
FAST
.
To display a ‘Stable’ reading more quickly and more often, use “
FAST
” setting.
MEDIUM
or
SLOW
is recommended for most applications. When Auto Read is enabled, a stability criteria setting of
“
SLOW
” is recommended to for best results.
SLOW:
The Stable indicator will take longer to appear and will appear less frequently. Use this
setting if you want the best results and don’t mind waiting longer to get them.
MEDIUM:
This is the factory default setting. It provides a balanced response which works best for
most applications.
FAST:
The Stable indicator will appear quickest of any setting. If you find that the Stable indicator
frequently appears and disappears, you may want to select a slower setting. This setting is not
recommended when
AUTO READ
is enabled. It is also not suitable for most non-refillable
electrodes which are generally slower to respond.
n
Navigation Of Tabs
SYSTEM SETUP & CONFIGURATION
PREV
NEXT
ESC
Scroll to previous
menu option
ESCAPE – Use to
cancel or exit out of
current screen
Scroll to next
menu option
SETUP
MEM IN
MEM OUT
REPORT
Enters Setup mode;
use to change
System settings or
measurement parameters
Access stored values
in memory
Inserts measured value
into memory
Access Calibration
Report Data
TEMP
SORT
CLR–C
NEXT–P
Access Temperature
Calibration
Clears current
calibration
Sorts stored data
(memory out) by
selected available
option
Skip to Next Page /
Next Parameter
of menu
MEAS
CAL
MODE
Resumes live
measurement
when Auto
Read is flashing
Toggles available
measurement modes
Begin Calibration
Mode using current
measured value
Sends data to
printer or PC
STABILITY
STABILITY CRITERIA
AUTO READ
BACKLIGHT
DATALOG
PRINT SETUP
PRINT FORMAT
PRINT SET
DATE & TIME
PASSWORD
CLEAR DATALOG
FACTORY RESET
CONTRAST
3
2