background image

5

4

1. Introduction

Thank you for purchasing an OAKTON

®

TEMP Acorn Meter. These economical,

microprocessor-based meters are available in three models:

• the Acorn TEMP 4 uses a series 400 thermistor temperature sensor

• the Acorn TEMP 5 uses a 100K thermistor temperature sensor

• the Acorn TEMP 6 uses a PT 100 RTD temperature sensor

All Acorn TEMP meters feature:

• Large LCD for clear and easy reading

• Readings in °C and °F (switchable)

• Minimum and maximum temperature display

• Minimum and maximum hold mode 

• Low battery indicator

• Hold function, freezes measured reading

• User calibration, offset adjustment

• Built-in memory backup; calibration and other information remain 

if battery is disconnected

• TEMP 5 thermistor sensor allows selection of most suitable 

temperature curve

This instruction manual is organized for easy reference. For basic functions of this
meter, read sections 2 through 5. These sections include basic instructions that will
get you up and running quickly. The remaining sections of this manual (6 through
11) deal with error messages and troubleshooting. This part of the manual also
includes the Specifications, Accessories, Warranty and Return of Items section. 

2. Display and Keypad Functions

The Acorn has a large custom LCD with the following indicators:

°

C

HO LO

 

°

F

MAX

MIN

Degree C

annunciator 

Degree F

annunciator 

minimum

Reading

indicates

HOLD

function is

activated

indicates low 

battery condition

maximum

The OAKTON Acorn meter has six keys on its splashproof keypad. These keys are
ON/OFF, CAL, HOLD/ENTER, MAX/

(DOWN), MIN/

(UP) and °C/°F.

ON/OFF:

Powers meter on and shuts 

unit off. Meter directly enters 
measurement mode when you 
turn it on.

CAL: 

Allows temperature calibration 

of the meter.

HOLD/ENTER: 

Freezes the measured 

reading; confirms calibration value.

MAX/

(UP) and MIN/

(DOWN): 

momentarily displays maximum 
or minimum temperature; 
enter maximum or minimum 
hold mode; scroll up and down 
in calibration mode.

°C/°F: 

Switches between °C and °F

in measurement mode.

ON

OFF

CAL

HOLD

ENTER

°

C/

°

F

MAX

MIN

Содержание Acorn TEMP 4

Страница 1: ...d please return freight prepaid and correction will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be r...

Страница 2: ...meter on 7 4 Measurement 8 9 4 1 Measuring temperature 8 4 2 Displaying minimum and maximum temperature 8 4 3 HOLD function 9 4 4 Minimum and maximum hold mode 9 5 Calibration 10 11 5 1 Temperature ca...

Страница 3: ...include basic instructions that will get you up and running quickly The remaining sections of this manual 6 through 11 deal with error messages and troubleshooting This part of the manual also includ...

Страница 4: ...ith a fresh set as soon as possible Connecting your Temperature sensor Insert the phono jack of the temperature sensor into the socket on the meter Unplug the phono jack when not in use 7 6 3 1 A A te...

Страница 5: ...deactivate the HOLD function and return to measurement mode Minimum and Maximum Hold Mode With Minimum and Maximum Hold Mode the Acorn TEMP meter can be used as a maximum registering or minimum regist...

Страница 6: ...LCD flashes CO the meter then reverts to temperature calibration mode and the display continues to flash See figures 7 Use the and keys to fine tune the temperature reading to your actual temperature...

Страница 7: ...l noises or Remove or switch off induction caused by interfering motor by nearby electric motor Dirty probe Clean probe Slow response Dirty probe Clean probe 7 Specifications Model TEMP 4 and 5 Thermi...

Страница 8: ...temperature probe Shpg wt 1 lb 460 g WD 35626 20 Acorn TEMP 6 RTD Thermometer with temperature probe Shpg wt 1 lb 460 g WD 35626 50 Replacement probe for TEMP 5 meter WD 35626 60 Replacement probe for...

Отзывы: