SET °C: Temperature settings
SET °C: Temperature settings
The temperature settings sub group lets you:
Toggle automatic temperature
compensation (ATC) on or off
If ATC selected on:
Temperature sensor calibration
If ATC selected off:
Set process temperature
Set calibration temperature
6.3
1
2
3
4
1
1
2
3
4
This set up group has two different sub paths. The screens you see
depend on if you select ATC on or off.
SETUP
HOLD
SETUP
HOLD
6.3.1 Selecting automatic or manual temperature compensation
From measurement mode:
1.
Enter Advanced set-up mode.
Push the ENTER key. Press the
▲
key to scroll to security code “22”.
Push the ENTER key again.
2.
Press the
▲
and
▼
keys
to scroll
until the upper display shows
SEt °C.
3.
Press the ENTER key.
The lower
display will show “Atc”; the upper
display will show “on” or “oFF”.
See figures at right
4.
Press the
▲
and
▼
keys
to toggle
between ATC on and off.
5.
Press the ENTER key
to confirm
selection.
6.
• If you selected ATC on:
Proceed to step 3 of section 6.3.2.
• If you selected ATC off:
Proceed to step 3 of section 6.3.3.
• To return to Measurement mode:
Press the
▲
and
▼
keys together.