Maintenance and storage
36
●
Maintenance and storage of pH and ORP electrodes
This section describes an overview of the procedures for maintenance and storage of pH
electrode.
●
How to clean the electrodes
When the tip of an electrode (responsive membrane and liquid junction) becomes dirty, the
response time may slow or an error may occur in the measurement results. To avoid such error,
clean the electrode. For dirt that cannot be washed off by pure water (or deionized water), use
the cleaning solution indicated below depending on the type of dirt. After cleaning, rinse the
electrode with pure water (or deionized water).
However, for pH and ORP electrodes, different cleaning solutions should be used to clean
different types of dirt.
For pH electrode
For ORP electrode
Type of dirt
Cleaning solution
General
Diluted neutral cleaning solution
Oil
Alcohol, or diluted neutral cleaning solution
Inorganic substance
1 mol/L HCl or electrode cleaning solution
Protein
Cleaning solution including protein-removing enzyme
Alkali
Dip in 1 mol/L HCl or electrode cleaning solution for 1h to 2 h
Type of dirt
Cleaning solution
General
Diluted neutral cleaning solution (General dish washing liquid
works reasonably well.)
Oil
Inorganic substance
Immerse dilute nitric acid (1:1 nitric acid)