55
Alarm Name,
Priority and
Type
Alarm Description
Solution
(High Priority)
(Technical)
seconds the screen will flash red at 2Hz,
sound a 10 beep sequence repeated
after 2.5 seconds of silence and the text
“No Flow” will flash in the system status
window
The unit will continue to sound until; the
power button is pressed to silence the
audio, the power button is held to power
off the unit, or 15 seconds has expired.
replace tubing or
humidifier. Clean or
replace air inlet filter.
Move device to ensure
adequate air flow.
Contact your authorized
dealer.
Over
Temperature
Alarm
(High Priority)
(Technical)
The maximum operating temperature of
the enclosure and pump is 70°C and
90°C respectively. When either of these
temperatures are reached the screen will
flash red at 2Hz, sound a 10 beep
sequence repeated after 2.5 seconds of
silence and the text “Over Temperature”
will flash in the system status window
The unit will continue to sound until; the
power button is pressed to silence the
audio, the power button is held to power
off the unit, or 15 seconds has expired.
Allow the device to
reach operating
temperature.
Utilize an alternate source
of oxygen if necessary.
Invalid Motor
Temperature
(High Priority)
(Technical)
When the motor temperature sensor is out
of a valid range the screen will flash red
at 2Hz, sound a 10 beep sequence
repeated after 2.5 seconds of silence and
the text “Invalid Motor Temperature” will
flash in the system status window
The unit will continue to sound until; the
power button is pressed to silence the
audio, the power button is held to power
off the unit, or 15 seconds has expired.
Allow the device to
reach operating
temperature.
Utilize an alternate source
of oxygen if necessary.
Invalid Box
Temperature
(High Priority)
(Technical)
When the internal box temperature
sensor is out of a valid range the screen
will flash red at 2Hz, sound a 10 beep
sequence repeated after 2.5 seconds of
silence and the text “Invalid Box
Temperature” will flash in the system
status window
The unit will continue to sound until; the
power button is pressed to silence the
audio, the power button is held to power
off the unit, or 15 seconds has expired.
Allow the device to
reach operating
temperature.
Utilize an alternate source
of oxygen if necessary.
Unauthorized
Battery Alert
(Medium
If the unit detects that a battery other
than one from O2 Concepts has been
installed in the unit the battery icon will
show an exclamation point within the
Remove unauthorized
battery and replace with
an authorized O2
Concepts battery.