Accessing the Alarm System Test
Use this test to verify that the alarm system is working properly. This test includes screen color, screen
flashing sequence, and audible alarm sequence. Follow the steps in the procedure listed below.
Press the Menu (M) button.
Press the Plus button (+) or Minus button (-) button to scroll to the Alarm Test icon.
Press the Menu button (M) to select Alarm Test.
The system will begin by testing the High Priority Alarm. The High Priority Alarm consists of a 10-beep
sequence separated by 2.5 seconds of silence and the LCD screen flashes red. The alarm symbol will
appear on the screen.
The device will automatically enter the Medium Priority Alarm test once the High Priority Alarm
sequence has concluded. Pressing the M button (M) while the High Priority Alarm test is in progress will
terminate the High Priority Alarm test and start the Medium Priority Alarm. The Medium Priority Alarm
consists of a 3-beep sequence separated by 15 seconds of silence and the LCD screen will flash
amber. The alarm symbol will appear on the screen.
The device will automatically enter the Low Priority Alarm test once the Medium Priority Alarm
sequence has concluded. Pressing the Menu button (M) while the Medium Priority Alarm test is in
progress will terminate the Medium Priority Alarm test and start the Low Priority Alarm. The Low Priority
Alarm consists of one (1) beep and a solid amber LCD screen. The alarm symbol will appear on the
screen.
At the conclusion of the test, the menu screen will appear. Use the Plus button (+) or Minus button (-) to
toggle to the Home screen. Press the Menu button (M) to return to the home screen.
The Remote Diagnostics and Cell Info screens are for use solely by O2 Concepts or your provider’s
technical support.