15
CAREvent® ALS - User Manual
Note:
If a single use patient circuit is used, safely dispose of after use.
The Resuscitator must be thoroughly cleaned after each patient use.
1. Operate CAREvent® ALS Handheld Resuscitator to blow out
any contaminant from the patient valve.
2. Ensure CAREvent® ALS Handheld Resuscitator is disconnected
from the gas supply source.
3. Remove the patient valve swivel housing (a) from the body of
the resuscitator (c), being careful to ensure that the diaphram
(b) is retained (fig 3).
4. Remove the facemask from the resuscitator (after removing the
mask retaining insert (if supplied) using the extraction tool).
5. Shake out any foreign material.
6. Wash all components thoroughly in a mild soap solution and
disinfect as required.
7. The resuscitator can be wiped over with a soft cloth and mild
soap solution.
8. Dry all components thoroughly.
9. Reassemble unit, connect to an air or oxygen supply to check
operation prior to packaging for emergency use.
5. CAREvent® ALS Handheld Resuscitator Accessories
ORDER N
0
PART
17MP9039
Reusable Patient Valve Swivel Housing
01PV1121
Reusable diaphragm (duckbill style)
01CV8015-Cs O-Two Deluxe Disposable Ventilation Circuit with PEEP Port
17MP1535-Cs
Anti-Entrainment Ring
01FG6500
Mask Retaining Insert
01FG6501
Mask Insert Extraction Tool
01FG6502
Head Harness System
01CV8010-Cs Single Use Transport Ventilation Circuit