Micro-Kill+™ (Micro-Kill Plus™) by Medline
(EPA Reg. No. 59894-10-37549)
To obtain disinfecting wipes and other information, please contact
Medline at 1-800-MEDLINE (1-800-633-5463) or visit www.medline.com.
Disinfecting Procedures
1.
Put on non-sterile gloves.
2.
Take out one disinfecting wipe from the package and squeeze out any
excess liquid in order to prevent damage to the meter.
3.
Wipe all exterior surface of the meter
including the display and buttons.
Hold the meter with the test strip slot
pointing down and wipe the area around
the test slot, be careful not to allow
excess liquid to get inside. Keep the
meter moist with disinfection solution
contained in the wipe for a minimum of
2 minutes for Micro-Kill+™ wipes. Follow
the instructions on the package label of
disinfecting wipe. Use two or more wipes
if necessary.
4.
Remove the wipe. Allow the meter surface to dry completely.
5.
Discard the used wipes and never reuse them.
6.
Remove and discard gloves in appropriate receptacles and wash hands.
Each cleaning and disinfection cycle includes a pre-cleaning step with
one wipe and a disinfection step with a second wipe.
This device has been validated to withstand 10,950 cleaning and